Recall Register

Device recall: Philips Xper Flex Cardio (FC2010, Procedure Room and FC2020, Bedside Monitor) devices - intended for use by…

Class II FDA enforcement event 87732. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

Performance issues with the Xper Flex Cardio Physio Monitoring System include: potential delay of up to 10 seconds in displaying ECG, invasive blood pressure and other parameters; patient weight is rounded to the nearest whole kilogram;Xper IM software used with the Xper Flex Cardio Physio Monitoring System may periodically crash; No SpO2 numeric or plethysmography is displayed when SpO2 is…

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Philips North America Llc started this device recall on 29 Mar 2021. The firm address in the FDA record is in Cambridge, MA. FDA put the recall in Class II. FDA event 87732 has 1 product record. The first FDA enforcement report date is 5 May 2021. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Nationwide Foreign: Argentina Australia Austria Bangladesh Belgium Bulgaria Canada Chile China Croatia Cyprus Czech Republic Denmark Dominican Republic Ecuador Egypt Estonia Finland France Germany…

Recalling firm

Philips North America Llc, Cambridge, MA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 87732: product records
Recall numberProductQuantityClassStatusReport date
Z-1517-2021Philips Xper Flex Cardio (FC2010, Procedure Room and FC2020, Bedside Monitor) devices - intended for use by professional healthcare providers for complete physiologic/hemodynamic monitoring Service numbers associated with FC2010: Xper Flex Cardio 2010 Rev A 453564241901 Xper Flex Cardio 2010 Rev B 453564483321 Flex…9208 units WWClass IIOngoing5 May 2021

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2021
900
Share of these events in Class II
89%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
129

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 87732. https://recallregister.com/fda/ii/87732/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Nationwide Foreign: Argentina Australia Austria Bangladesh Belgium Bulgaria Canada Chile China Croatia Cyprus Czech Republic Denmark Dominican Republic Ecuador Egypt Estonia Finland France Germany…
What quantity of product is in FDA event 87732?
The FDA record gives this quantity: 9208 units WW
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2021 have the same class?
89% of the 900 device events that started in 2021 are in Class II.