Device recall: Implantable Pulse Generators - Product Usage: used to treat arrhythmias related to bradycardia. Additionally, CRT-P…

Class II FDA enforcement event 87662. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

A longevity estimation error may occur in the early years of device life when a unipolar pacing vector is programmed in the right atrial (RA) lead and/or the right ventricular (RV) lead. No other device features or therapies are impacted.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) started this device recall on 6 Apr 2021. The firm address in the FDA record is in Mounds View, MN. FDA put the recall in Class II. FDA event 87662 has 4 product records. The first FDA enforcement report date is 12 May 2021. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide distribution.

Recalling firm

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF), Mounds View, MN.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

4 product records in the event.

FDA event 87662: product records
Recall numberProductQuantityClassStatusReport date
Z-1525-2021Implantable Pulse Generators - Product Usage: used to treat arrhythmias related to bradycardia. Additionally, CRT-P devices can also provide cardiac resynchronization therapy (CRT) in treatment of heart failure. 1. Azure family: Azure XT DR MRI SureScan", Azure S DR MRI SureScan, Azure XT SR MRI SureScan, Azure S SR…575815 unitsClass IIOngoing12 May 2021
Z-1526-2021Implantable cardiac resynchronization therapy pacemakers (CRT-P) - Product Usage: used to treat arrhythmias related to bradycardia. Additionally, CRT-P devices can also provide cardiac resynchronization therapy (CRT) in treatment of heart failure. 1. Percepta family: Percepta CRT-P MRI SureScan, Percepta Quad CRT-P…333408 unitsClass IIOngoing12 May 2021
Z-1527-2021Programmers - Product Usage: used to treat arrhythmias related to bradycardia. Additionally, CRT-P devices can also provide cardiac resynchronization therapy (CRT) in treatment of heart failure. 1. 2090 Carelink 2. 29901 Encore37690 unitsClass IIOngoing12 May 2021
Z-1528-2021CareLink SmartSync Device Manager - Product Usage: used to treat arrhythmias related to bradycardia. Additionally, CRT-P devices can also provide cardiac resynchronization therapy (CRT) in treatment of heart failure, Model Number 24970A16,311 unitsClass IIOngoing12 May 2021

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2021
900
Share of these events in Class II
89%
Product records in this event
4Median of the firm: 2
Events of the firm on the register
30

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 87662. https://recallregister.com/fda/ii/87662/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide distribution.
What quantity of product is in FDA event 87662?
The event has 4 product records. The first record gives this quantity: 575815 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2021 have the same class?
89% of the 900 device events that started in 2021 are in Class II.