Recall Register

Device recall: Randox Lipase Assay: Lipase Colorimetric Reagent - Product Usage: A Lipase test system is a device intended for the…

Class II FDA enforcement event 86039. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Randox have confirmed imprecision of quality control and patient samples when using specific Randox Lipase R2 Substrate batches.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

RANDOX LABORATORIES, LTD. started this device recall on 26 Jun 2020. The firm address in the FDA record is in Crumlin Colorado Antrim. FDA put the recall in Class II. FDA event 86039 has 3 product records. The first FDA enforcement report date is 12 Aug 2020. The record gives the status Terminated, with the termination date 19 Oct 2020.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide distribution - US Nationwide distribution including in the states of Puerto Rico and WV.

Recalling firm

RANDOX LABORATORIES, LTD., Crumlin Colorado Antrim, United Kingdom.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

3 product records in the event.

FDA event 86039: product records
Recall numberProductQuantityClassStatusReport date
Z-2751-2020Randox Lipase Assay: Lipase Colorimetric Reagent - Product Usage: A Lipase test system is a device intended for the quantitative in vitro determination of Lipase in human serum and plasma.2219 kitsClass IITerminated12 Aug 2020
Z-2752-2020Randox Lipase Assay: Lipase Colorimetric R1 Buffer - Product Usage: A Lipase test system is a device intended for the quantitative in vitro determination of Lipase in human serum and plasma.Class IITerminated12 Aug 2020
Z-2753-2020Randox Lipase Assay: Lipase Colorimetric R2 Substrate - Product Usage: A Lipase test system is a device intended for the quantitative in vitro determination of Lipase in human serum and plasma.Class IITerminated12 Aug 2020

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2020
916
Share of these events in Class II
93%
Product records in this event
3Median of the firm: 1
Events of the firm on the register
52

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 86039. https://recallregister.com/fda/ii/86039/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide distribution - US Nationwide distribution including in the states of Puerto Rico and WV.
What quantity of product is in FDA event 86039?
The event has 3 product records. The first record gives this quantity: 2219 kits
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2020 have the same class?
93% of the 916 device events that started in 2020 are in Class II.