Food, drugs, devices › Randox Laboratories Ltd.
Device recall: RX Series Copper (Cu) Assay Ref. Number CU2340
Randox Laboratories Ltd. (Crumlin (North)) initiated this device recall on 29 Nov 2023; FDA lists 1 product record under event 93775, classified Class II.
FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 88% of the 917 device enforcement events initiated in 2023 carry the same class.
Status: Ongoing. Initiation: voluntary: firm initiated. Randox Laboratories Ltd. has 52 enforcement events on this register, with a median of 1 product record per event.
Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.
Recalling firm
Randox Laboratories Ltd., Crumlin (North), Ireland.
Reason for recall
"Randox Laboratories have released an update to the Carryover Avoidance Technical Bulletin to detail that Copper, should not be run directly after Total Protein, on the RX series instruments testing order. If copper assay is run after total protein it could lead to falsely elevated copper results."
Distribution
"US Nationwide distribution in the states of SC and OK."
Product records
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-0992-2024 | RX Series Copper (Cu) Assay Ref. Number CU2340 | 3,053 kits (2 US, 3,051 OUS) | Class II | Ongoing | 14 Feb 2024 |
Context
Other enforcement events for Randox Laboratories Ltd.
All 52 events for Randox Laboratories Ltd.
Questions and answers
Why was this device recalled?
FDA's record gives the reason for recall as: "Randox Laboratories have released an update to the Carryover Avoidance Technical Bulletin to detail that Copper, should not be run directly after Total Protein, on the RX series instruments testing order. If copper assay is run after total protein it could lead to falsely elevated copper results."
What does Class II mean?
FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.
Where was the product distributed?
FDA's distribution pattern field reads: "US Nationwide distribution in the states of SC and OK."
Is the recall still ongoing?
FDA lists the status as Ongoing. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.
Where does this data come from?
openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.