Food, drugs, devices › NuVasive Inc
Device recall: Nuvasive NVM5 EMG Module, REF 8020015, Rx ONLY, CE2797, Nonsterile, UDI: 00887517206633 - Product Usage: is intended for
FDA enforcement event 85889 is a device recall by NuVasive Inc of San Diego, CA, initiated 10 Jun 2020 and classified Class II, with 1 product record.
FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 93% of the 916 device enforcement events initiated in 2020 carry the same class.
Status: Terminated, terminated 22 Mar 2021. Initiation: voluntary: firm initiated. NuVasive Inc has 6 enforcement events on this register, with a median of 1 product record per event.
Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.
Recalling firm
NuVasive Inc, San Diego, CA.
Reason for recall
"The NVM5 EMG Module kit intended for intraoperative neurophysiologic monitoring during spinal surgery contains incorrect component, part number 1715945, SSEP harness instead of the the correct part number 1747065, NVM5 Harness, Gen 3.5 EM. This issue can result in delay in therapy or lack of EMG neuromonitoring during surgery if the surgeon elects to proceed without the NVM5."
Distribution
"US Nationwide distribution including in the states of AR, CA, IA, IL, WI. OUS: None"
Product records
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-2684-2020 | Nuvasive NVM5 EMG Module, REF 8020015, Rx ONLY, CE2797, Nonsterile, UDI: 00887517206633 - Product Usage: is intended for intraoperative neurophysiologic monitoring during spinal surgery. | 23 | Class II | Terminated | 29 Jul 2020 |
Context
Other enforcement events for NuVasive Inc
| Initiated | Product (first record) | Type | Class | Records |
|---|---|---|---|---|
| 10 Nov 2022 | X-CORE 2 Ti Core, Static. Part of the NuVasive X-Core Expandable VBR System used with spinal fixation. | Device | Class II | 1 |
| 19 Jul 2022 | NuVasive MD Pulse III Multimodality System REF 1881500 Pulse Software version 369-BL.US and 369-BLN.US | Device | Class III | 1 |
| 20 Jun 2022 | Description/Part: Simplify Disc Size 1, Ht 5/SM-5; Simplify Disc Size 1, Ht 6/SM-6; Simplify Disc Size 2, Ht 4/MD-4; Sim | Device | Class II | 1 |
| 13 Feb 2020 | NUVASIVE MAGEC 2 Rod X.X mm XX mm, REF MC2-XXXXS LOT 0000000 AAA - Product Usage: comprised of a sterile single-use spin | Device | Class II | 1 |
| 5 Jan 2009 | Screw Tulip , 6.25mm Rod Subsequent Codes: KWP, KWQ, MNH, MNI | Device | Class II | 31 |
Questions and answers
Why was this device recalled?
FDA's record gives the reason for recall as: "The NVM5 EMG Module kit intended for intraoperative neurophysiologic monitoring during spinal surgery contains incorrect component, part number 1715945, SSEP harness instead of the the correct part number 1747065, NVM5 Harness, Gen 3.5 EM. This issue can result in delay in therapy or lack of EMG neuromonitoring during surgery if the surgeon elects to proceed without the NVM5."
What does Class II mean?
FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.
Where was the product distributed?
FDA's distribution pattern field reads: "US Nationwide distribution including in the states of AR, CA, IA, IL, WI. OUS: None"
Is the recall still ongoing?
FDA lists the status as Terminated; the termination date is 22 Mar 2021. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.
Where does this data come from?
openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.
Provenance
Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.