Device recall: NUVASIVE MAGEC 2 Rod X.X mm XX mm, REF MC2-XXXXS LOT 0000000 AAA - Product Usage: comprised of a sterile single-use spin

FDA enforcement event 85495 · Class II · initiated 13 Feb 2020 · status Terminated · terminated 25 May 2022

DeviceVoluntary: Firm initiated1 product recordSan Diego, CA

NuVasive Inc (San Diego, CA) initiated this device recall on 13 Feb 2020; FDA lists 1 product record under event 85495, classified Class II.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 93% of the 916 device enforcement events initiated in 2020 carry the same class.

Status: Terminated, terminated 25 May 2022. Initiation: voluntary: firm initiated. NuVasive Inc has 6 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

NuVasive Inc, San Diego, CA.

Reason for recall

"Post-implantation separation of an actuator end cap component that may expose internal components of the actuator, which could lead to hastened degeneration of the internal components and egress of Titanium alloy wear debris and resultant localized tissue discoloration."

Distribution

"Worldwide distribution - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NY,…"

Product records

Recall numberProductQuantityClassStatusReport date
Z-1898-2020NUVASIVE MAGEC 2 Rod X.X mm XX mm, REF MC2-XXXXS LOT 0000000 AAA - Product Usage: comprised of a sterile single-use spinal rod that is surgically implanted using appropriate NuVasive Reline, Reline Small Stature, and Armada fixation components (i.e. Pedicle screws, hooks and/or connectors, provided non-sterile, to be…3502 rodsClass IITerminated13 May 2020

Context

93%of 2020 device events are Class II916 events that year
1product record in this eventfirm median: 1
6events on this register for the firm

Other enforcement events for NuVasive Inc

InitiatedProduct (first record)TypeClassRecords
10 Nov 2022X-CORE 2 Ti Core, Static. Part of the NuVasive X-Core Expandable VBR System used with spinal fixation.DeviceClass II1
19 Jul 2022NuVasive MD Pulse III Multimodality System REF 1881500 Pulse Software version 369-BL.US and 369-BLN.USDeviceClass III1
20 Jun 2022Description/Part: Simplify Disc Size 1, Ht 5/SM-5; Simplify Disc Size 1, Ht 6/SM-6; Simplify Disc Size 2, Ht 4/MD-4; SimDeviceClass II1
10 Jun 2020Nuvasive NVM5 EMG Module, REF 8020015, Rx ONLY, CE2797, Nonsterile, UDI: 00887517206633 - Product Usage: is intended forDeviceClass II1
5 Jan 2009Screw Tulip , 6.25mm Rod Subsequent Codes: KWP, KWQ, MNH, MNIDeviceClass II31

All 6 events for NuVasive Inc

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "Post-implantation separation of an actuator end cap component that may expose internal components of the actuator, which could lead to hastened degeneration of the internal components and egress of Titanium alloy wear debris and resultant localized tissue discoloration."

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "Worldwide distribution - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NY,…"

Is the recall still ongoing?

FDA lists the status as Terminated; the termination date is 25 May 2022. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.