Recall Register

Device recall: HeartStart MRx Monitor/Defibrillators with model numbers M3535A M3536A, M3536M, M3536MC, M3536M2, M3536M3, M3536M4…

Class II FDA enforcement event 84841. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Philips has received a number of reports of HeartStart MRx Monitor/Defibrillators that have suffered internal damage and were not able to deliver therapy after having been dropped or subjected to a severe mechanical shock, even though the device did not have visible external damage or the Ready for Use ( RFU ) indicator on the unit did not immediately indicate a problem. One of these reports…

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Philips North America, LLC started this device recall on 26 Feb 2020. The firm address in the FDA record is in Andover, MA. FDA put the recall in Class II. FDA event 84841 has 1 product record. The first FDA enforcement report date is 25 Mar 2020. The record gives the status Terminated, with the termination date 8 Dec 2022.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

US Nationwide domestic distribution. Worldwide foreign distribution.

Recalling firm

Philips North America, LLC, Andover, MA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 84841: product records
Recall numberProductQuantityClassStatusReport date
Z-1499-2020HeartStart MRx Monitor/Defibrillators with model numbers M3535A M3536A, M3536M, M3536MC, M3536M2, M3536M3, M3536M4, M3536M5, M3536M6, M3536M7, M3536M8, and M3536M9.97788 unitsClass IITerminated25 Mar 2020

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2020
916
Share of these events in Class II
93%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
129

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 84841. https://recallregister.com/fda/ii/84841/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: US Nationwide domestic distribution. Worldwide foreign distribution.
What quantity of product is in FDA event 84841?
The FDA record gives this quantity: 97788 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2020 have the same class?