Recall Register

Device recall: HEMO-7 Hemospray Endoscopic Hemostat

Class II FDA enforcement event 84822. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Complaints that the handle and/or activation knob on Hemostat devices have cracked or broken when the device is activated. In some cases, this has caused the carbon dioxide cartridge to exit the handle. Most of these complaints reported that the carbon dioxide cartridge exited the handle with minimal force. However, some complaints reported that the carbon dioxide cartridge exited the handle…

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Wilson-Cook Medical Inc. started this device recall on 4 Feb 2020. The firm address in the FDA record is in Winston Salem, NC. FDA put the recall in Class II. FDA event 84822 has 2 product records. The first FDA enforcement report date is 29 Apr 2020. The record gives the status Terminated, with the termination date 15 Jan 2021.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Nationwide distribution. International distribution to Bahrain, Belarus, Canary Islands, Gibraltar, Guadeloupe, Guernsey, Iraq, Israel, Jersey, Jordan, Kuwait, Lebanon, Martinique, Monaco, Morocco,…

Recalling firm

Wilson-Cook Medical Inc., Winston Salem, NC.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

2 product records in the event.

FDA event 84822: product records
Recall numberProductQuantityClassStatusReport date
Z-1745-2020HEMO-7 Hemospray Endoscopic Hemostat89,423 devices totalClass IITerminated29 Apr 2020
Z-1746-2020HEMO-10 Hemospray Endoscopic Hemostat89,423 devices totalClass IITerminated29 Apr 2020

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2020
916
Share of these events in Class II
93%
Product records in this event
2Median of the firm: 1
Events of the firm on the register
14

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 84822. https://recallregister.com/fda/ii/84822/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Nationwide distribution. International distribution to Bahrain, Belarus, Canary Islands, Gibraltar, Guadeloupe, Guernsey, Iraq, Israel, Jersey, Jordan, Kuwait, Lebanon, Martinique, Monaco, Morocco,…
What quantity of product is in FDA event 84822?
The event has 2 product records. The first record gives this quantity: 89,423 devices total
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2020 have the same class?
93% of the 916 device events that started in 2020 are in Class II.