Recall Register

Food, drugs, devices › Wilson-Cook Medical Inc.

Device recall: QUANTUM TTC BILIARY BALLOON DILATOR, REF QBD-10X3, Order Number G22657; used to dilate strictures of the biliary tree

FDA enforcement event 93326 · Class II · initiated 3 Nov 2023 · status Ongoing

DeviceVoluntary: Firm initiated7 product recordsWinston Salem, NC

Wilson-Cook Medical Inc. (Winston Salem, NC) initiated this device recall on 3 Nov 2023; FDA lists 7 product records under event 93326, classified Class II.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 88% of the 917 device enforcement events initiated in 2023 carry the same class.

Status: Ongoing. Initiation: voluntary: firm initiated. Wilson-Cook Medical Inc. has 14 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Wilson-Cook Medical Inc., Winston Salem, NC.

Reason for recall

"Nonconforming devices while manufactured correctly, do not comply to their design requirements. Potential harms associated with balloon detachment are endoscopic retrieval of an impacted or non-impacted object, bleeding caused by the retrieval device, or the incident can occur without injury and the object is left to pass naturally."

Distribution

"US"

Product records

Recall numberProductQuantityClassStatusReport date
Z-0457-2024QUANTUM TTC BILIARY BALLOON DILATOR, REF QBD-10X3, Order Number G22657; used to dilate strictures of the biliary tree731 devicesClass IIOngoing13 Dec 2023
Z-0458-2024QUANTUM TTC BILIARY BALLOON DILATOR, REF QBD-10X3-E, Order Number G22667; used to dilate strictures of the biliary tree111 devicesClass IIOngoing13 Dec 2023
Z-0459-2024QUANTUM TTC BILIARY BALLOON DILATOR, REF QBD-4X3, Order Number G22654; used to dilate strictures of the biliary tree553 devicesClass IIOngoing13 Dec 2023
Z-0460-2024QUANTUM TTC BILIARY BALLOON DILATOR, REF QBD-6X3, Order Number G22655; used to dilate strictures of the biliary tree941 devicesClass IIOngoing13 Dec 2023
Z-0461-2024QUANTUM TTC BILIARY BALLOON DILATOR, REF QBD-6X3-E, Order Number G22765; used to dilate strictures of the biliary tree218 devicesClass IIOngoing13 Dec 2023
Z-0462-2024QUANTUM TTC BILIARY BALLOON DILATOR, REF QBD-8X3, Order Number G22656; used to dilate strictures of the biliary tree888 devicesClass IIOngoing13 Dec 2023
Z-0463-2024QUANTUM TTC BILIARY BALLOON DILATOR, REF QBD-8X3-E, Order Number G22766; used to dilate strictures of the biliary tree157 devicesClass IIOngoing13 Dec 2023

Context

88%of 2023 device events are Class II917 events that year
7product records in this eventfirm median: 1
14events on this register for the firm

Other enforcement events for Wilson-Cook Medical Inc.

InitiatedProduct (first record)TypeClassRecords
20 Aug 2026Brand Name: HEMOSPRAY ENDOSCOPIC HEMOSTAT Product Name: HEMOSPRAY ENDOSCOPIC HEMOSTAT Model/Catalog Number: HEMO-7, HEMODeviceClass II1
3 Mar 2026Instinct Plus Endoscopic Clipping Device REF: G58010 Rx Only, Sterile EODeviceClass II1
18 Dec 2025Teslatome Bipolar Sphincterotomes REF: TESLA-B2535-260 Model Number: G60784 UDI-DI code: 00827002607842 REF: TESLA-B2535DeviceClass II1
15 Dec 2025Product Name: Nasal Jejunal Feeding Tube REF: NJFT-10DeviceClass II1
31 Oct 2024Cook Medical HEMO-7-EU Hemospray Endoscopic Hemostat, REF G24663, 2.8 cm, sterile.DeviceClass II1
29 Aug 2024Cook Medical Hemo-7 Hemospray Endoscopic Hemostat, REF G56572, 2.8 mm, Sterile; and Hemo-10 Hemospray Endoscopic HemostaDeviceClass II1
22 Jun 2023Fusion Lithotripsy Extraction Basket, Model Number FS-LXB-2X4 G48277; Fusion Lithotripsy Extraction Baskets are intendedDeviceClass II2
22 May 2023Hemospray Endoscopic Hemostat: Reference Part Numbers: a) HEMO-7, REF G56572; b) HEMO-7-EU, REF G24663;DeviceClass II2
4 Feb 2020HEMO-7 Hemospray Endoscopic HemostatDeviceClass II2
24 Jul 2019Hemospray Endoscopic HemostatDeviceClass II1

All 14 events for Wilson-Cook Medical Inc.

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "Nonconforming devices while manufactured correctly, do not comply to their design requirements. Potential harms associated with balloon detachment are endoscopic retrieval of an impacted or non-impacted object, bleeding caused by the retrieval device, or the incident can occur without injury and the object is left to pass naturally."

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "US"

Is the recall still ongoing?

FDA lists the status as Ongoing. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.