Recall Register

Device recall: Arrow Berman Angiographic Catheter 50cc Item Number: AI-07130

Class II FDA enforcement event 84399. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Potential leak in the balloon extension line. If a leak was present and went undetected during pre-use testing as outlined in the IFU, this would cause the balloon to deflate slowly following placement. If the leak went undetected by attending clinicians during use, it could result in contrast injection into the vascular or cardiac tissue which could result in serious injury or death

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Arrow International Inc started this device recall on 15 Nov 2019. The firm address in the FDA record is in Reading, PA. FDA put the recall in Class II. FDA event 84399 has 2 product records. The first FDA enforcement report date is 25 Dec 2019. The record gives the status Terminated, with the termination date 19 Jun 2020.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

U.S. Nationwide, Canada

Recalling firm

Arrow International Inc, Reading, PA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

2 product records in the event.

FDA event 84399: product records
Recall numberProductQuantityClassStatusReport date
Z-0715-2020Arrow Berman Angiographic Catheter 50cc Item Number: AI-0713071 unitsClass IITerminated25 Dec 2019
Z-0716-2020Arrow Berman Angiographic Catheter 60cm Item Number: AI-07130-J1 unitClass IITerminated25 Dec 2019

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2019
968
Share of these events in Class II
88%
Product records in this event
2Median of the firm: 1
Events of the firm on the register
93

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 84399. https://recallregister.com/fda/ii/84399/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: U.S. Nationwide, Canada
What quantity of product is in FDA event 84399?
The event has 2 product records. The first record gives this quantity: 71 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2019 have the same class?
88% of the 968 device events that started in 2019 are in Class II.