Recall Register

Food, drugs, devices › ARROW INTERNATIONAL Inc.

Device recall: ARROW QuickFlash Radial Artery Catherization (wire, guide, catheter), Part Numbers: a) REF AK-04220; b) REF ASK-04220-FM

FDA enforcement event 94033 · Class I · initiated 12 Feb 2024 · status Ongoing

DeviceVoluntary: Firm initiated1 product recordMorrisville, NC

ARROW INTERNATIONAL Inc. (Morrisville, NC) initiated this device recall on 12 Feb 2024; FDA lists 1 product record under event 94033, classified Class I.

FDA defines Class I as a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death. 11% of the 970 device enforcement events initiated in 2024 carry the same class.

Status: Ongoing. Initiation: voluntary: firm initiated. ARROW INTERNATIONAL Inc. has 93 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

ARROW INTERNATIONAL Inc., Morrisville, NC.

Reason for recall

"Teleflex is initiating this voluntary recall for the above-mentioned products due to reports received indicating a potentially defective component. The complaints received relate to resistance of the guidewire handle/chamber during use. The possible immediate health consequences of the component issue are arterial vasospasm and vessel injury arising from multiple arterial punctures with repeated…"

Distribution

"Worldwide distribution: US (nationwide) and OUS (foreign) to countries of: AE, AR, AT, AU, BE, BH, BR, CA, CN, CO, CY, DE, EC, EE, ES, FR, GR, HK, HR, IT, KW, LB, LU, MX, MY, NL, NZ, PH, PL, QA, RU,…"

Product records

Recall numberProductQuantityClassStatusReport date
Z-1267-2024ARROW QuickFlash Radial Artery Catherization (wire, guide, catheter), Part Numbers: a) REF AK-04220; b) REF ASK-04220-FMH1; c) REF ASK-04220-HHC; d) REF ASK-04220-KSP; e) REF ASK-04220-PSU; f) REF ASK-04220-UCL1; g) REF ASK-04500-AH; h) REF ASK-04500-HF-S; i) REF NA-04220-S1A; j) REF NA-04220-X1A; k) REF RA-04220; l)…1,160,470 devicesClass IOngoing27 Mar 2024

Context

11%of 2024 device events are Class I970 events that year
1product record in this eventfirm median: 1
93events on this register for the firm

Other enforcement events for ARROW INTERNATIONAL Inc.

InitiatedProduct (first record)TypeClassRecords
26 Jul 2024Arrow Arterial Kits and Trays as follows: (1) Arrow Arterial Access Tray with 0.025" Spring-Wire Guide (Catheter Not IncDeviceClass II1
29 Apr 2024Arrow FiberOptix Intra-Aortic Balloon Catheter Kit: a) REF IAB-05830-LWS; b) REF IAB-05840-LWS; c) REF IAB-05850-LWSDeviceClass I3
10 Aug 2023Arrow ErgoPack, Pressure Injectable Arrowg+ard Blue Advance Two-Lumen PICC pre-loaded with Arrow VPS Precision Stylet, RDeviceClass I4
10 Aug 2023Arrow ErgoPack Pressure Injectable One-Lumen PICC, REF CDC-05541-HPK1A; catheter, intravascularDeviceClass II2
19 May 2023ARROW Endurance Extended Dwell Peripheral Catheter SystemDeviceClass I1
27 Oct 2022Arrow MAC Two-Lumen Central Venous Access Kit fr use with 7 - 7.5 Fr Catheters, REF ASK-11242-UPM1; catheter introducerDeviceClass I2
17 Oct 2022Arrow AutoCAT2 Intra-Aortic Balloon Pump, AUTOCAT 2 IAPB, REF IAP-0400 (IPN000302), cardiac pumpDeviceClass I31
14 Sep 2022ARROW Pressure Injectable Arrowg+ard Blue Advance Two-Lumen Midline Kit, 5.5 Fr, 15 cm, .018 Inch dia., REF ASK-41552-JHDeviceClass II1
5 May 2022Arrow Three-Lumen CVCDeviceClass II1
20 Sep 2021Arrow-Trerotola Over-The-Wire PTD Kit Percutaneous Thrombolytic Device : 7FR Product Code: PT-12709-WC The Arrow-TrerotoDeviceClass I4

All 93 events for ARROW INTERNATIONAL Inc.

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "Teleflex is initiating this voluntary recall for the above-mentioned products due to reports received indicating a potentially defective component. The complaints received relate to resistance of the guidewire handle/chamber during use. The possible immediate health consequences of the component issue are arterial vasospasm and vessel injury arising from multiple arterial punctures with repeated…"

What does Class I mean?

FDA classifies recalls by relative degree of health hazard. Class I is a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death.

Where was the product distributed?

FDA's distribution pattern field reads: "Worldwide distribution: US (nationwide) and OUS (foreign) to countries of: AE, AR, AT, AU, BE, BH, BR, CA, CN, CO, CY, DE, EC, EE, ES, FR, GR, HK, HR, IT, KW, LB, LU, MX, MY, NL, NZ, PH, PL, QA, RU,…"

Is the recall still ongoing?