Device recall: MultiDiagnost-Eleva, Product codes 708032, 708036 Product Usage: As a multifunctional I universal imaging application…

Class II FDA enforcement event 82313. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

The locking plate may be missing from the system, which may result in unexpected stand/table tilt movement that cannot be stopped by the user.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Philips North America, LLC started this device recall on 28 Feb 2019. The firm address in the FDA record is in Andover, MA. FDA put the recall in Class II. FDA event 82313 has 5 product records. The first FDA enforcement report date is 3 Apr 2019. The record gives the status Terminated, with the termination date 3 Aug 2020.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

US Nationwide domestic distribution, and worldwide international distribution.

Recalling firm

Philips North America, LLC, Andover, MA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

5 product records in the event.

FDA event 82313: product records
Recall numberProductQuantityClassStatusReport date
Z-1052-2019MultiDiagnost-Eleva, Product codes 708032, 708036 Product Usage: As a multifunctional I universal imaging application system, General RIF, Fluoroscopy, Radiography and Angiography can be performed along with more specialized interventional applications on human patients. This includes the following general areas:…1969Class IITerminated3 Apr 2019
Z-1053-2019Urodiagnost, Product No. 708033 Product Usage: The Philips UroDiagnost MRF can be used to perform all urological and gynecological routine examinations including: Urograms with injection or infusion Retrograde pyelography Cystography and urethrography Micturition cystography Cystoscopy, endopscopy Transurethral…1969Class IITerminated3 Apr 2019
Z-1054-2019Omnidiagnost Classic, Product Codes 70859, 708023, 708024, 708025 Product Usage: The Philips OmniDiagnost is a multi-functional, universal X-ray system with remote-controlled facilities. It performs a wide range of routine examinations and interventional procedures include: Gastro-intestinal examination Urography…1696Class IITerminated3 Apr 2019
Z-1055-2019MultiDiagnost Eleva w/ Flat Detector, Product codes 708034, 708037, 708038 Product Usage: As a multifunctional I universal imaging application system, General RIF, Fluoroscopy, Radiography and Angiography can be performed along with more specialized interventional applications on human patients. This includes the…1969Class IITerminated3 Apr 2019
Z-1056-2019URODiagnost Eleva, Product codes 708026, 708027, 708028 Product Usage: The Philips OmniDiagnost is a multi-functional, universal X-ray system with remote-controlled facilities. It performs a wide range of routine examinations and interventional procedures include: Gastro-intestinal examination Urography Bronchography…1969Class IITerminated3 Apr 2019

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2019
968
Share of these events in Class II
88%
Product records in this event
5Median of the firm: 1
Events of the firm on the register
129

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 82313. https://recallregister.com/fda/ii/82313/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: US Nationwide domestic distribution, and worldwide international distribution.
What quantity of product is in FDA event 82313?
The event has 5 product records. The first record gives this quantity: 1969
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2019 have the same class?
88% of the 968 device events that started in 2019 are in Class II.