Recall Register

Food, drugs, devices › Howmedica Osteonics Corp.

Device recall: Tritanium TL Steerable Inserter, Cat. No. 48930000, UDI 07613327381900 Product Usage: The TL Steerable Inserter is a Cla

FDA enforcement event 81096 · Class II · initiated 12 Sep 2018 · status Terminated · terminated 16 Jul 2020

DeviceVoluntary: Firm initiated1 product recordAllendale, NJ

FDA enforcement event 81096 is a device recall by Howmedica Osteonics Corp. of Allendale, NJ, initiated 12 Sep 2018 and classified Class II, with 1 product record.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 92% of the 1,074 device enforcement events initiated in 2018 carry the same class.

Status: Terminated, terminated 16 Jul 2020. Initiation: voluntary: firm initiated. Howmedica Osteonics Corp. has 48 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Howmedica Osteonics Corp., Allendale, NJ.

Reason for recall

"The firm received complaints related to the locking mechanism on the steerable inserter not functioning as intended. The firm confirmed that the issue is related to lack of lubrication of the instrument. There are two potential hazards associated with the implant pivoting unexpectedly during implant insertion into the disc space. First, if the implant is initially inserted into the disc space…"

Distribution

"US Nationwide Distribution."

Product records

Recall numberProductQuantityClassStatusReport date
Z-0207-2019Tritanium TL Steerable Inserter, Cat. No. 48930000, UDI 07613327381900 Product Usage: The TL Steerable Inserter is a Class 1 instrument which is provided in a non-sterile condition. The inserter is designed to enter the disc space, with a fixed engagement between the inserter and the cage and then rotate for desired…58Class IITerminated31 Oct 2018

Context

92%of 2018 device events are Class II1,074 events that year
1product record in this eventfirm median: 1
48events on this register for the firm

Other enforcement events for Howmedica Osteonics Corp.

InitiatedProduct (first record)TypeClassRecords
30 Jul 2026AXLE PIN ASSY LARGE ELBOW COBALT CHROME Catalog No. C-M112-2-500 Orthopaedic implant components manufactured from UHMWPEDeviceClass II12
19 May 2026TRI TS BASEPLATE SIZE 3. Part Number: 5521-B-300.DeviceClass II2
2 Feb 2026MICS3 Angled Sagittal Saw Attachment; Part Number: 210490DeviceClass II1
12 Nov 2025The Stryker CranialMask Tracker is a single-use device consisting of a flexible printed circuit board (PCB) with 31 infrDeviceClass II1
28 Aug 20251. EXETER V40 STEM 44MM NO 2, Model/Catalog Number: 0580-1-442; 2. EXETER V40 STEM 37.5MM NO 0, Model/Catalog Number: 05DeviceClass II1
4 Sep 2024Triathlon Universal TS Baseplate Size 2 - A single use sterile implantable device intended for cemented applications durDeviceClass II1
7 Aug 2024MAKO Integrated Cutting System (MICS) HANDPIECE, Catalog Number 209063DeviceClass II1
5 Aug 2024Custom Made Device - GMRS EXTENSION Piece 40MM LEFT, 35 Deg EXT Rotated Tabs (Patient Specific); Catalog Number: C-MM100DeviceClass II1
19 Jun 2024UNITRAX Endoprosthesis Head Component - 43mm. Implantable component used in partial hip arthroplasty. Catalog Number: 69DeviceClass II3
23 May 2024HRIS Storage & Sterilization Case and Tray, Catalog/Part Number: 6210-9-900DeviceClass II2

All 48 events for Howmedica Osteonics Corp.

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "The firm received complaints related to the locking mechanism on the steerable inserter not functioning as intended. The firm confirmed that the issue is related to lack of lubrication of the instrument. There are two potential hazards associated with the implant pivoting unexpectedly during implant insertion into the disc space. First, if the implant is initially inserted into the disc space…"

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "US Nationwide Distribution."

Is the recall still ongoing?

FDA lists the status as Terminated; the termination date is 16 Jul 2020. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.