Recall Register

Device recall: Visualase Cooled Laser Applicator System (VCLAS) labeled as: a. Part Number 9735559, Visualase Cooled Laser Applicator…

Class II FDA enforcement event 80481. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Medtronic has become aware that certain lots of the VCLAS 3MM, 10MM, and 15MM laser diffusing fibers were not tested at intended laser power levels, due to an issue related to calibration of the test equipment, which can lead to the potential for low laser energy output from fiber or fiber burnout at lower wattage.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Medtronic Navigation, Inc. started this device recall on 23 Mar 2018. The firm address in the FDA record is in Louisville, CO. FDA put the recall in Class II. FDA event 80481 has 1 product record. The first FDA enforcement report date is 25 Jul 2018. The record gives the status Terminated, with the termination date 31 Jul 2023.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution - US Nationwide in the states of AL, AR, AZ, CA, CO, DC, FL, GA, IL, IN, KY, LA, MA, MD, MI, MN, MO, NC, NJ, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI. International…

Recalling firm

Medtronic Navigation, Inc., Louisville, CO.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 80481: product records
Recall numberProductQuantityClassStatusReport date
Z-2501-2018Visualase Cooled Laser Applicator System (VCLAS) labeled as: a. Part Number 9735559, Visualase Cooled Laser Applicator System 3mm Tip; b. Part Number 9735560 Visualase Cooled Laser Applicator System 10mm Tip; c. Part Number 9735561 Visualase Cooled Laser Applicator System 15mm Tip Product Usage: The Visualase…1,283 devices totalClass IITerminated25 Jul 2018

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2018
1,074
Share of these events in Class II
92%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
44

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 80481. https://recallregister.com/fda/ii/80481/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution - US Nationwide in the states of AL, AR, AZ, CA, CO, DC, FL, GA, IL, IN, KY, LA, MA, MD, MI, MN, MO, NC, NJ, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI. International…
What quantity of product is in FDA event 80481?
The FDA record gives this quantity: 1,283 devices total
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2018 have the same class?