Device recall: StealthStation S8 App versions 1.3.0 and 1.3.2, (Model # 9735762) The StealthStation System, with StealthStation Spine s

FDA enforcement event 95283 · Class II · initiated 22 Aug 2024 · status Ongoing

DeviceVoluntary: Firm initiated1 product recordLafayette, CO

Medtronic Navigation, Inc. (Lafayette, CO) initiated this device recall on 22 Aug 2024; FDA lists 1 product record under event 95283, classified Class II.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 87% of the 970 device enforcement events initiated in 2024 carry the same class.

Status: Ongoing. Initiation: voluntary: firm initiated. Medtronic Navigation, Inc. has 44 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Medtronic Navigation, Inc., Lafayette, CO.

Reason for recall

"Due to two software anomalies that can occur within a spine procedure. Two issues are: 1. Navigation Orientation - when lateral orientation radiographic images are taken, information may appear flipped when utilizing the system (180 degree flipped). 2. Pop-up Message - when using CT+Fluouroscopy images with a slice spacing thickness greater than 2mm, the system does not display a pop-up message…"

Distribution

"Worldwide - U.S Nationwide distribution including in the states of AK, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NH, NM, NV, NY, OH, OK, PA, Puerto Rico,…"

Product records

Recall numberProductQuantityClassStatusReport date
Z-0050-2025StealthStation S8 App versions 1.3.0 and 1.3.2, (Model # 9735762) The StealthStation System, with StealthStation Spine software, is intended as an aid for precisely locating anatomical structures in either open or percutaneous neurosurgical and orthopedic procedures.1,801 systemsClass IIOngoing16 Oct 2024

Context

87%of 2024 device events are Class II970 events that year
1product record in this eventfirm median: 1
44events on this register for the firm

Other enforcement events for Medtronic Navigation, Inc.

InitiatedProduct (first record)TypeClassRecords
4 Dec 2024Medtronic PIN, 100mm STERILE PERCUTANEOUS REF: 9733235 Medtronic PIN, 150MM, STERILE PERCUTANEOUS REF: 97733236 The SterDeviceClass II1
26 Mar 2024Stealth S8 Clinical Software Application, REF: 9735762, CE0344, Rx Only. Used in the StealthStation System as an aid forDeviceClass I1
28 Nov 2023Medtronic, REF 9733235, Pin, 100mm, STERILE, PERCUTANEOUSDeviceClass II2
27 Sep 2023StealthStation S8 Application version 2.0 and 2.0.1 (Part 9735762) The StealthStation System, with StealthStation CraniaDeviceClass I1
11 Apr 2023StealthStation Cranial Software, Model: 9735585, used with StealthStation S7/i7 systemsDeviceClass I1
11 Apr 2023StealthStation Cranial Software, Models: 9735585, 9735586 (kit), 9735587 (kit), used with StealthStation S7/i7 systemsDeviceClass I1
15 Dec 2022AxiEM" Non-Invasive Patient TrackerDeviceClass II1
8 Dec 2022RadiaLux Lighted Retractor (Pink), REF: 50-101-1DeviceClass II1
14 Oct 2022Visualase Cooled Laser Applicator System (VCLAS) 9735560 .4MM CORE FIBER 10MM TIP product. Used to necrotize or coagulatDeviceClass II1
11 Nov 2021Stealthstation System w/ Stealthstation Cranial Software 3.1.1,3.1.2,3.1. or SynergyCranial Software. Synergy Cranial, MDeviceClass I1

All 44 events for Medtronic Navigation, Inc.

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "Due to two software anomalies that can occur within a spine procedure. Two issues are: 1. Navigation Orientation - when lateral orientation radiographic images are taken, information may appear flipped when utilizing the system (180 degree flipped). 2. Pop-up Message - when using CT+Fluouroscopy images with a slice spacing thickness greater than 2mm, the system does not display a pop-up message…"

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "Worldwide - U.S Nationwide distribution including in the states of AK, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NH, NM, NV, NY, OH, OK, PA, Puerto Rico,…"

Is the recall still ongoing?

FDA lists the status as Ongoing. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.