Recall Register

Device recall: Stealthstation System w/ Stealthstation Cranial Software 3.1.1,3.1.2,3.1. or SynergyCranial Software. Synergy Cranial…

Class I FDA enforcement event 89058. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

Cranial biopsy procedure software can enter a state where the biopsy depth gauge is no longer synchronized with the rest of the navigational information on the screen and may display an incorrect position of the biopsy needle, which may result in prolonged procedure, the need for additional surgical procedure, tissue injury.

Class I, in the words of FDA

a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death

Medtronic Navigation, Inc. started this device recall on 11 Nov 2021. The firm address in the FDA record is in Louisville, CO. FDA put the recall in Class I. FDA event 89058 has 1 product record. The first FDA enforcement report date is 29 Dec 2021. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

US: MN, FL, MD, IL, AK, NE, NH, PA, ND, AR, CA, WI, NJ, GA, SD, AZ, KY, AL, MS, MO, TX, MI, MT, OH, NY, VA, MA, CT, TN, WV, WA, NC, CO, NV, ME, LA, DE, NM, IN, UT, DC, OR, SC, OK, IA, HI, ID, KS,…

Recalling firm

Medtronic Navigation, Inc., Louisville, CO.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 89058: product records
Recall numberProductQuantityClassStatusReport date
Z-0383-2022Stealthstation System w/ Stealthstation Cranial Software 3.1.1,3.1.2,3.1. or SynergyCranial Software. Synergy Cranial, Model: 9733763, and StealthStation Cranial, Model: 97355852587Class IOngoing29 Dec 2021

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class I: The highest of the three FDA recall classes.

The event in context

Device events that started in 2021
900
Share of these events in Class I
8%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
44

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 89058. https://recallregister.com/fda/i/89058/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: US: MN, FL, MD, IL, AK, NE, NH, PA, ND, AR, CA, WI, NJ, GA, SD, AZ, KY, AL, MS, MO, TX, MI, MT, OH, NY, VA, MA, CT, TN, WV, WA, NC, CO, NV, ME, LA, DE, NM, IN, UT, DC, OR, SC, OK, IA, HI, ID, KS,…
What quantity of product is in FDA event 89058?
The FDA record gives this quantity: 2587
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2021 have the same class?
8% of the 900 device events that started in 2021 are in Class I.