Recall Register

Food, drugs, devices › Medtronic Navigation, Inc.

Device recall: Visualase Thermal Therapy System, Software: Visualase Software versions 3.1.1 - 3.2, Product #: 9735542

FDA enforcement event 80308 · Class II · initiated 1 Jun 2018 · status Terminated · terminated 2 Feb 2022

DeviceVoluntary: Firm initiated1 product recordLouisville, CO

Medtronic Navigation, Inc. (Louisville, CO) initiated this device recall on 1 Jun 2018; FDA lists 1 product record under event 80308, classified Class II.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 92% of the 1,074 device enforcement events initiated in 2018 carry the same class.

Status: Terminated, terminated 2 Feb 2022. Initiation: voluntary: firm initiated. Medtronic Navigation, Inc. has 44 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Medtronic Navigation, Inc., Louisville, CO.

Reason for recall

"Inaccuracy of MR thermometry during MRI-guided laser ablation procedures using the Medtronic Visualase Thermal Therapy System may result in unaccounted spread of thermal energy to the surrounding tissue."

Distribution

"U.S.: MN, NY, PA, FL, AZ, MA, VA, CA, CO, AL, WI, TX, MO, DC, OH, NH, GA, SC, NJ, WA, CT, MI, OK, IN, MD, IL, KY, LA, OR, RI, UT, AR, MS, TN, NC Foreign: United Kingdom, Germany, Switzerland,…"

Product records

Recall numberProductQuantityClassStatusReport date
Z-3251-2018Visualase Thermal Therapy System, Software: Visualase Software versions 3.1.1 - 3.2, Product #: 9735542161Class IITerminated3 Oct 2018

Context

92%of 2018 device events are Class II1,074 events that year
1product record in this eventfirm median: 1
44events on this register for the firm

Other enforcement events for Medtronic Navigation, Inc.

InitiatedProduct (first record)TypeClassRecords
4 Dec 2024Medtronic PIN, 100mm STERILE PERCUTANEOUS REF: 9733235 Medtronic PIN, 150MM, STERILE PERCUTANEOUS REF: 97733236 The SterDeviceClass II1
22 Aug 2024StealthStation S8 App versions 1.3.0 and 1.3.2, (Model # 9735762) The StealthStation System, with StealthStation Spine sDeviceClass II1
26 Mar 2024Stealth S8 Clinical Software Application, REF: 9735762, CE0344, Rx Only. Used in the StealthStation System as an aid forDeviceClass I1
28 Nov 2023Medtronic, REF 9733235, Pin, 100mm, STERILE, PERCUTANEOUSDeviceClass II2
27 Sep 2023StealthStation S8 Application version 2.0 and 2.0.1 (Part 9735762) The StealthStation System, with StealthStation CraniaDeviceClass I1
11 Apr 2023StealthStation Cranial Software, Model: 9735585, used with StealthStation S7/i7 systemsDeviceClass I1
11 Apr 2023StealthStation Cranial Software, Models: 9735585, 9735586 (kit), 9735587 (kit), used with StealthStation S7/i7 systemsDeviceClass I1
15 Dec 2022AxiEM" Non-Invasive Patient TrackerDeviceClass II1
8 Dec 2022RadiaLux Lighted Retractor (Pink), REF: 50-101-1DeviceClass II1
14 Oct 2022Visualase Cooled Laser Applicator System (VCLAS) 9735560 .4MM CORE FIBER 10MM TIP product. Used to necrotize or coagulatDeviceClass II1

All 44 events for Medtronic Navigation, Inc.

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "Inaccuracy of MR thermometry during MRI-guided laser ablation procedures using the Medtronic Visualase Thermal Therapy System may result in unaccounted spread of thermal energy to the surrounding tissue."

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "U.S.: MN, NY, PA, FL, AZ, MA, VA, CA, CO, AL, WI, TX, MO, DC, OH, NH, GA, SC, NJ, WA, CT, MI, OK, IN, MD, IL, KY, LA, OR, RI, UT, AR, MS, TN, NC Foreign: United Kingdom, Germany, Switzerland,…"

Is the recall still ongoing?

FDA lists the status as Terminated; the termination date is 2 Feb 2022. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.