Food, drugs, devices › Ortho-Clinical Diagnostics
Device recall: VITROS Chemistry Products PHYT Slides, REF/CATALOG 829 8671, Rx ONLY, IVD; Ortho-Clinical Diagnostics, Inc. For in vitro
FDA enforcement event 77613 is a device recall by Ortho-Clinical Diagnostics of Rochester, NY, initiated 12 Jun 2017 and classified Class II, with 1 product record.
FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 93% of the 1,091 device enforcement events initiated in 2017 carry the same class.
Status: Terminated, terminated 11 Aug 2020. Initiation: voluntary: firm initiated. Ortho-Clinical Diagnostics has 105 enforcement events on this register, with a median of 1 product record per event.
Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.
Recalling firm
Ortho-Clinical Diagnostics, Rochester, NY.
Reason for recall
"The firm identified the potential for biased results to be generated when processing VITROS PHYT Slide product with citrate plasma specimen samples. The current Instructions for Use (IFU) and Assay Summary Chart state serum, heparin plasma, and citrate plasma as acceptable specimen types for use with the PHYT assay."
Distribution
"Worldwide Distribution - US Distribution including Puerto Rico and the following countries: Bermuda, Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan,…"
Product records
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-0273-2018 | VITROS Chemistry Products PHYT Slides, REF/CATALOG 829 8671, Rx ONLY, IVD; Ortho-Clinical Diagnostics, Inc. For in vitro diagnostic use only. VITROS Chemistry Products PHYT Slides quantitatively measure phenytoin (PHYT) concentration in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and… | US; 4680; Foreign: 1286 | Class II | Terminated | 3 Jan 2018 |
Context
Other enforcement events for Ortho-Clinical Diagnostics
All 105 events for Ortho-Clinical Diagnostics
Questions and answers
Why was this device recalled?
FDA's record gives the reason for recall as: "The firm identified the potential for biased results to be generated when processing VITROS PHYT Slide product with citrate plasma specimen samples. The current Instructions for Use (IFU) and Assay Summary Chart state serum, heparin plasma, and citrate plasma as acceptable specimen types for use with the PHYT assay."
What does Class II mean?
FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.
Where was the product distributed?
FDA's distribution pattern field reads: "Worldwide Distribution - US Distribution including Puerto Rico and the following countries: Bermuda, Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan,…"
Is the recall still ongoing?
FDA lists the status as Terminated; the termination date is 11 Aug 2020. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.