Recall Register

Device recall: BrightView X, model #882478; gamma camera designed for single or dual detector nuclear imaging

Class II FDA enforcement event 74894. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

The firm became aware of an issue where during a non-Auto Body Contouring (ABC), non-circular, rel-180 scan, the detector may come in contact with the patient. The issue was caused by a software defect in the workflow setup script.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Philips Medical Systems (Cleveland) Inc started this device recall on 25 May 2016. The firm address in the FDA record is in Cleveland, OH. FDA put the recall in Class II. FDA event 74894 has 3 product records. The first FDA enforcement report date is 14 Sep 2016. The record gives the status Terminated, with the termination date 2 Jul 2018.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide distribution. US nationwide, Algeria, Argentina, Armenia, Australia, Austria, Belgium, Brazil, Czech Republic, Denmark, Finland France, Gabon, Germany, Greece, India, Indonesia, Ireland,…

Recalling firm

Philips Medical Systems (Cleveland) Inc, Cleveland, OH.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

3 product records in the event.

FDA event 74894: product records
Recall numberProductQuantityClassStatusReport date
Z-2735-2016BrightView X, model #882478; gamma camera designed for single or dual detector nuclear imaging117 unitsClass IITerminated14 Sep 2016
Z-2736-2016BrightView XCT Model 882482, gamma camera for Single Photon Emission Computed Tomography (SPECT)424 unitsClass IITerminated14 Sep 2016
Z-2737-2016BrightView XCT Upgrade Model 8824543 unitsClass IITerminated14 Sep 2016

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2016
1,232
Share of these events in Class II
91%
Product records in this event
3Median of the firm: 1
Events of the firm on the register
129

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 74894. https://recallregister.com/fda/ii/74894/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide distribution. US nationwide, Algeria, Argentina, Armenia, Australia, Austria, Belgium, Brazil, Czech Republic, Denmark, Finland France, Gabon, Germany, Greece, India, Indonesia, Ireland,…
What quantity of product is in FDA event 74894?
The event has 3 product records. The first record gives this quantity: 117 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2016 have the same class?
91% of the 1,232 device events that started in 2016 are in Class II.