Device recall: Brilliance 64 CT Model number 728231 Product Usage: Computed Tomography X-ray systems intended to produce…
Class II FDA enforcement event 74820. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that is not planned at iso-center (0,0). If the region of interest is not visualized in the images, a CT rescan may be performed.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Philips Medical Systems (Cleveland) Inc started this device recall on 20 Jul 2016. The firm address in the FDA record is in Cleveland, OH. FDA put the recall in Class II. FDA event 74820 has 7 product records. The first FDA enforcement report date is 31 Aug 2016. The record gives the status Terminated, with the termination date 2 Jul 2018. 54 of the 1,232 device events that started in 2016 have more product records. The median device event of 2016 has 1 product record. The openFDA file has 1,232 device events that started in 2016 and 1,153 that started in 2015.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
US Nationwide Distribution in the states of: AL,AR,AZ,CA,CO,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA, MD,ME,MI,MN,MO,MS,NC,ND,NE,NJ,NM, NV,NY,OH,OK,OR,PA,PR,SC,TN,TX,UT,VA,VT,WA,WI,WV,WY
Recalling firm
Philips Medical Systems (Cleveland) Inc, Cleveland, OH.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
7 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-2652-2016 | Brilliance 64 CT Model number 728231 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. | 283 | Class II | Terminated | 31 Aug 2016 |
| Z-2653-2016 | Brilliance CT Big Bore Oncology, Model number 728243 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. | 33 | Class II | Terminated | 31 Aug 2016 |
| Z-2654-2016 | Brilliance BigBore Radiology CT Model number 728244 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. | 24 | Class II | Terminated | 31 Aug 2016 |
| Z-2655-2016 | Brilliance 16, Model number 728246 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. | 192 | Class II | Terminated | 31 Aug 2016 |
| Z-2656-2016 | Brilliance CT 16 Power, Model number 728240 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. | 50 | Class II | Terminated | 31 Aug 2016 |
| Z-2657-2016 | Ingenuity Core, Model number 728321 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. | 30 | Class II | Terminated | 31 Aug 2016 |
| Z-2658-2016 | Ingenuity Core128, Model number 728323 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. | 3 | Class II | Terminated | 31 Aug 2016 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2016
- 1,232
- Share of these events in Class II
- 91%
- Product records in this event
- 7Median of the firm: 1
- Events of the firm on the register
- 129
Other events of the firm
Official channels
- FDA enforcement report Z-2652-2016The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 74820. https://recallregister.com/fda/ii/74820/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 10 Nov 2026
- Newest FDA report date on the register
- 4 Nov 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
US Nationwide Distribution in the states of: AL,AR,AZ,CA,CO,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA, MD,ME,MI,MN,MO,MS,NC,ND,NE,NJ,NM, NV,NY,OH,OK,OR,PA,PR,SC,TN,TX,UT,VA,VT,WA,WI,WV,WY
- What quantity of product is in FDA event 74820?
- The event has 7 product records. The first record gives this quantity:
283
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2016 have the same class?
- 91% of the 1,232 device events that started in 2016 are in Class II.