Recall Register

Device recall: Torque Limiting Handle: PRECIMED, T-Handle Torque, Non Sterile. Handle: Customer Number/Model: MCL120014 T14000…

Class II FDA enforcement event 73948. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Greatbatch Medical has initiated a global Field Recall for Torque Limiting Devices. Greatbatch identified through laboratory testing that the required sterility assurance level (SAL) of 10-6 is not achieved on torque limiting devices when subjected to steam sterilization cycles identified in the Instruction For Use (IFU) provided with the device.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Greatbatch Medical started this device recall on 26 Apr 2016. The firm address in the FDA record is in Plymouth, MN. FDA put the recall in Class II. FDA event 73948 has 3 product records. The first FDA enforcement report date is 8 Jun 2016. The record gives the status Terminated, with the termination date 7 Jun 2018.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

US: IN, MI, TX. OUS: GERMANY, FRANCE, ITALY, SWITZERLAND.

Recalling firm

Greatbatch Medical, Plymouth, MN.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

3 product records in the event.

FDA event 73948: product records
Recall numberProductQuantityClassStatusReport date
Z-1882-2016Torque Limiting Handle: PRECIMED, T-Handle Torque, Non Sterile. Handle: Customer Number/Model: MCL120014 T14000, TLT911-50-S20 TLT911-50-S20, TLT911-15-S01 TLT911-15-S01, TLTS0007-20-S19 TLTS0007-20-S19, TLT911-35-S01 TLT911-35-S01.2Class IITerminated8 Jun 2016
Z-1883-2016Torque Limiting Screwdriver: Greatbatch Medical, Torque Limiting Driver, Non Sterile, Rx Only. PRECIMED, Non Sterile. Zimmer, Non Sterile, Rx Only. Screwdriver: Customer Number/Model: 804-03-038 T11623, T14001 T14001, MTO35173507 T14378, MTO375173506 T14379, MTO35173505 T14380, MTO35175303 T8127.1110Class IITerminated8 Jun 2016
Z-1884-2016Torque Limiting Adapter: Torque Limiting Attachment, Non Sterile, Distributed by Zimmer. PRECIMED, Torque Limiting Adaptor, Non Sterile.. Customer Number/Model: 00-2360-080-00 T10232, T116245 T116245, TLA025-25 TLA025-25, 00-2360-080-05 T10233, MAC21003 T116430, TLA025-30 TLA025-30, MAC21001 T112972, MAC21001 T13621,…5952Class IITerminated8 Jun 2016

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2016
1,232
Share of these events in Class II
91%
Product records in this event
3Median of the firm: 1
Events of the firm on the register
14

Other events of the firm

Other FDA enforcement events of Greatbatch Medical
InitiatedProduct (first record)TypeClassRecords
23 Apr 2017Linear Straight Broach Handle (Rasp Handle) (Greatbatch Part # 510397; DJO Surgical Inc., # 803-03-159) and EZ Clean" LiDeviceClass II1
15 Feb 2017Biosense Webster MobiCath Bi-Directional Guiding Sheath, Model D140010 (small curve) and D140011 (large curve). The MobiDeviceClass II1
11 Nov 2016ViaPeel PTFE Peelable Introducer 5FR, Part Number 10890-006DeviceClass II1
19 Oct 2016Offset Reamer Handle, T5766 Product Usage: The reamer handles are re-usable surgical instruments used in arthroplasty suDeviceClass II1
8 Sep 2016Greatbatch Stiffer Coaxial Micro-Introducer. Model #/ Name: 10904-001 Stiffer Coaxial Micro-Introducer 4FR; 10904-002 StDeviceClass II1
2 Aug 2016Greatbatch, Offset Cup Impactor with the following Parts and Catalogue #: Tl0600 ~ MMP3100CHA04 T10491 MIMP3100CHA04 T12DeviceClass I1
28 Apr 2014Coaxial Micro-Introducer Kit. Greatbatch Part Number/Bard Access Systems Order Number/Part Description 10636-001 8004011DeviceClass II1
31 Jan 2014Greatbatch Medical, Offset Cup Impactor. Part Numbers: P9485, T10243, T10281, T10287, T10484, T10491, T10507, T10600, T1DeviceClass I1
31 May 2013Cordis a Johnson & Johnson Company, RadialSource Spring Wire Transradial Access Kit. Manufacturer Greatbatch Medical 230DeviceClass II1
16 Jan 201345¿ Angle Driver / Z-H, Greatbatch Medical 10 000 Wahrie Drive, Clarance, NY 14031. Model #: P2661, P2662, T3635, T5586,DeviceClass II1

Greatbatch Medical: all 14 events

Official channels

Cite this page

Recall Register. FDA enforcement event 73948. https://recallregister.com/fda/ii/73948/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: US: IN, MI, TX. OUS: GERMANY, FRANCE, ITALY, SWITZERLAND.
What quantity of product is in FDA event 73948?
The event has 3 product records. The first record gives this quantity: 2
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2016 have the same class?
91% of the 1,232 device events that started in 2016 are in Class II.