Recall Register

Device recall: Greatbatch Medical, Offset Cup Impactor. Part Numbers: P9485, T10243, T10281, T10287, T10484, T10491, T10507, T10600…

Class I FDA enforcement event 67392. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Greatbatch Medical has initiated a global Field Corrective Action for the Standard Offset Cup Impactor since the product was first released in 2004. Greatbatch has identified through internal Sterility Assurance Level validation testing that a sterility assurance level of 10-6 is not achieved on the Standard Offset Cup Impactor when it is subjected to the steam sterilization cycles and…

Class I, in the words of FDA

a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death

Greatbatch Medical started this device recall on 31 Jan 2014. The firm address in the FDA record is in Minneapolis, MN. FDA put the recall in Class I. FDA event 67392 has 1 product record. The first FDA enforcement report date is 12 Mar 2014. The record gives the status Terminated, with the termination date 31 Oct 2016.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution - USA including CA, IN, KS, MA, NJ, TN,TX,UT and Internationally to Australia, Austria, Belgium, China, France, Germany, Greece, Ireland, Italy, Japan, Luxembourg,…

Recalling firm

Greatbatch Medical, Minneapolis, MN.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 67392: product records
Recall numberProductQuantityClassStatusReport date
Z-1119-2014Greatbatch Medical, Offset Cup Impactor. Part Numbers: P9485, T10243, T10281, T10287, T10484, T10491, T10507, T10600, T10604, T10753, T10861, T11209, T11340, T12121, T12230, T12288, T12305, T12391, T12699, T12767, T12861, T13344, T13480, T13642, T13722, T13999, T14155, T14384, T15311, T15752, T15822, T15949, T16117,…4870 worldwide (2482 US), EXPANDED: 7245 devicesClass ITerminated12 Mar 2014

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class I: The highest of the three FDA recall classes.

The event in context

Device events that started in 2014
1,176
Share of these events in Class I
5%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
14

Other events of the firm

Other FDA enforcement events of Greatbatch Medical
InitiatedProduct (first record)TypeClassRecords
23 Apr 2017Linear Straight Broach Handle (Rasp Handle) (Greatbatch Part # 510397; DJO Surgical Inc., # 803-03-159) and EZ Clean" LiDeviceClass II1
15 Feb 2017Biosense Webster MobiCath Bi-Directional Guiding Sheath, Model D140010 (small curve) and D140011 (large curve). The MobiDeviceClass II1
11 Nov 2016ViaPeel PTFE Peelable Introducer 5FR, Part Number 10890-006DeviceClass II1
19 Oct 2016Offset Reamer Handle, T5766 Product Usage: The reamer handles are re-usable surgical instruments used in arthroplasty suDeviceClass II1
8 Sep 2016Greatbatch Stiffer Coaxial Micro-Introducer. Model #/ Name: 10904-001 Stiffer Coaxial Micro-Introducer 4FR; 10904-002 StDeviceClass II1
2 Aug 2016Greatbatch, Offset Cup Impactor with the following Parts and Catalogue #: Tl0600 ~ MMP3100CHA04 T10491 MIMP3100CHA04 T12DeviceClass I1
26 Apr 2016Torque Limiting Handle: PRECIMED, T-Handle Torque, Non Sterile. Handle: Customer Number/Model: MCL120014 T14000, TLT911-DeviceClass II3
28 Apr 2014Coaxial Micro-Introducer Kit. Greatbatch Part Number/Bard Access Systems Order Number/Part Description 10636-001 8004011DeviceClass II1
31 May 2013Cordis a Johnson & Johnson Company, RadialSource Spring Wire Transradial Access Kit. Manufacturer Greatbatch Medical 230DeviceClass II1
16 Jan 201345¿ Angle Driver / Z-H, Greatbatch Medical 10 000 Wahrie Drive, Clarance, NY 14031. Model #: P2661, P2662, T3635, T5586,DeviceClass II1

Greatbatch Medical: all 14 events

Official channels

Cite this page

Recall Register. FDA enforcement event 67392. https://recallregister.com/fda/i/67392/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution - USA including CA, IN, KS, MA, NJ, TN,TX,UT and Internationally to Australia, Austria, Belgium, China, France, Germany, Greece, Ireland, Italy, Japan, Luxembourg,…
What quantity of product is in FDA event 67392?
The FDA record gives this quantity: 4870 worldwide (2482 US), EXPANDED: 7245 devices
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2014 have the same class?
5% of the 1,176 device events that started in 2014 are in Class I.