Recall Register

Device recall: BD Trucount Absolute Counting Tubes: Model number: 340334 IVD; BD Trucount tubes are used for determining absolute…

Class II FDA enforcement event 73748. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

One lot of lot of BD Trucount tubes may contain tubes from another lot. There is a 3.3% difference in bead counts between the two lots found within some of the pouches and presents a low risk to patient safety or impact to data if this product is used.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Becton, Dickinson and Company, BD Biosciences started this device recall on 31 Mar 2016. The firm address in the FDA record is in San Jose, CA. FDA put the recall in Class II. FDA event 73748 has 1 product record. The first FDA enforcement report date is 4 May 2016. The record gives the status Terminated, with the termination date 26 May 2017.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution - US including MD, CA GA. WA. PA. CO. TN. FL. NC. DE. VA, CT, LA. NC, NY, OH, MN, NH and Internationally to China, Japan, and India.

Recalling firm

Becton, Dickinson and Company, BD Biosciences, San Jose, CA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 73748: product records
Recall numberProductQuantityClassStatusReport date
Z-1570-2016BD Trucount Absolute Counting Tubes: Model number: 340334 IVD; BD Trucount tubes are used for determining absolute counts of leucocytes in blood. BD Trucount tubes are designed for use with in vitro diagnostic products such as BD Tritest reagents, and a suitably equipped flow cytometer. BD Trucount tubes can be used…1743; 467 US and 1276 InternationalClass IITerminated4 May 2016

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2016
1,232
Share of these events in Class II
91%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
22

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 73748. https://recallregister.com/fda/ii/73748/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution - US including MD, CA GA. WA. PA. CO. TN. FL. NC. DE. VA, CT, LA. NC, NY, OH, MN, NH and Internationally to China, Japan, and India.
What quantity of product is in FDA event 73748?
The FDA record gives this quantity: 1743; 467 US and 1276 International
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2016 have the same class?
91% of the 1,232 device events that started in 2016 are in Class II.