Device recall: BD Trucount Tubes (Cat. No. 340334), used for determining absolute counts of leucocytes in blood. BD Trucount tubes…
Class II FDA enforcement event 90866. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
Due to partial label detachment on tubes. This can cause tubes to become stuck together in the pouch or cause automation errors when used with the automated loader for the wash assistant. Firm root cause investigation confirmed that label detachment is due to an adhesive change by a tier 2 supplier.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Becton, Dickinson and Company, BD Biosciences started this device recall on 3 Oct 2022. The firm address in the FDA record is in San Jose, CA. FDA put the recall in Class II. FDA event 90866 has 3 product records. The first FDA enforcement report date is 9 Nov 2022. The record gives the status Ongoing.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
U.S.: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SD, TN, TX, UT, VA, VT, WA, and WI O.U.S.:…
Recalling firm
Becton, Dickinson and Company, BD Biosciences, San Jose, CA.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
3 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-0171-2023 | BD Trucount Tubes (Cat. No. 340334), used for determining absolute counts of leucocytes in blood. BD Trucount tubes are designed for use with in vitro diagnostic products such as BD Tritest reagents, and a suitably equipped flow cytometer. BD Trucount tubes can be used with the BD FACS Loader. | 52,570 tubes (corrected on 11/04/2022) | Class II | Ongoing | 9 Nov 2022 |
| Z-0172-2023 | BD Trucount Tubes (Cat. No. 663028), used for determining absolute counts of leukocytes in blood. BD Trucount Tubes are designed for use with in vitro diagnostic products such as BD Tritest reagents, and a suitably equipped flow cytometer. BD Trucount Tubes can be used with the BD FACS Loader. | 11,709 tubes (corrected 11/04/2022) | Class II | Ongoing | 9 Nov 2022 |
| Z-0173-2023 | BD Multitest 6-Color TBNK with optional BD Trucount Tubes (Cat. No. 662995), for use with BD FACSLyric, BD FACSCanto II, and BD FACSCanto flow cytometers to determine the percentages and absolute counts of human lymphocyte subsets in peripheral whole blood for immunophenotyping. | 1,457 tubes (corrected on 11/04/2022) | Class II | Ongoing | 9 Nov 2022 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2022
- 839
- Share of these events in Class II
- 88%
- Product records in this event
- 3Median of the firm: 1
- Events of the firm on the register
- 22
Other events of the firm
Becton, Dickinson and Company, BD Biosciences: all 22 events
Official channels
- FDA enforcement report Z-0171-2023The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 90866. https://recallregister.com/fda/ii/90866/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
U.S.: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SD, TN, TX, UT, VA, VT, WA, and WI O.U.S.:…
- What quantity of product is in FDA event 90866?
- The event has 3 product records. The first record gives this quantity:
52,570 tubes (corrected on 11/04/2022)
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2022 have the same class?
- 88% of the 839 device events that started in 2022 are in Class II.