Recall Register

Device recall: Brilliance 64 Computed Tomography X-ray system

Class II FDA enforcement event 72684. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Perfusion scan feature may not be available on machines running software versions 4.1.2, 4.1.3, and 4.1.4. Customers previously using the basic axial perfusion software (versions preceding 4.1.2) without buying the helical perfusion license key could not access the basic axial perfusion feature after upgrading to software version 4.1.2.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Philips Medical Systems (Cleveland) Inc started this device recall on 29 Oct 2015. The firm address in the FDA record is in Cleveland, OH. FDA put the recall in Class II. FDA event 72684 has 6 product records. The first FDA enforcement report date is 30 Dec 2015. The record gives the status Terminated, with the termination date 26 Jun 2018.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide distribution. US nationwide, Australia, Austria, Bangladesh, Belgium, Brazil, Cambodia, Canada, China, Egypt, France, Germany, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Korea,…

Recalling firm

Philips Medical Systems (Cleveland) Inc, Cleveland, OH.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

6 product records in the event.

FDA event 72684: product records
Recall numberProductQuantityClassStatusReport date
Z-0544-2016Brilliance 64 Computed Tomography X-ray system30 unitsClass IITerminated30 Dec 2015
Z-0545-2016Brilliance iCT Computed Tomography X-ray system51 UnitsClass IITerminated30 Dec 2015
Z-0546-2016Brilliance iCT SP Computed Tomography X-ray system12 UnitsClass IITerminated30 Dec 2015
Z-0547-2016Ingenuity Core Computed Tomography X-ray system135 UnitsClass IITerminated30 Dec 2015
Z-0548-2016Ingenuity Core 128 Computed Tomography X-ray system177 UnitsClass IITerminated30 Dec 2015
Z-0549-2016Ingenuity CT Computed Tomography X-ray system31 UnitsClass IITerminated30 Dec 2015

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2015
1,153
Share of these events in Class II
91%
Product records in this event
6Median of the firm: 1
Events of the firm on the register
129

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 72684. https://recallregister.com/fda/ii/72684/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide distribution. US nationwide, Australia, Austria, Bangladesh, Belgium, Brazil, Cambodia, Canada, China, Egypt, France, Germany, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Korea,…
What quantity of product is in FDA event 72684?
The event has 6 product records. The first record gives this quantity: 30 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2015 have the same class?
91% of the 1,153 device events that started in 2015 are in Class II.