Recall Register

Device recall: Medtronic Passive Biopsy Needle Kit, Part Number 9733068 - Passive Biopsy Needle Kit. The image guided Passive Biopsy…

Class II FDA enforcement event 71756. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

The adhesive between the sphere assembly and the biopsy needle shaft may be compromised, causing the sphere assembly to move on the needle shaft. This may result in inaccurate navigation and improper placement of the needles biopsy window in the brain.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Medtronic Navigation, Inc. started this device recall on 17 Jul 2015. The firm address in the FDA record is in Louisville, CO. FDA put the recall in Class II. FDA event 71756 has 2 product records. The first FDA enforcement report date is 12 Aug 2015. The record gives the status Terminated, with the termination date 7 Apr 2016.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution - US (nationwide) and Canada.

Recalling firm

Medtronic Navigation, Inc., Louisville, CO.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

2 product records in the event.

FDA event 71756: product records
Recall numberProductQuantityClassStatusReport date
Z-2324-2015Medtronic Passive Biopsy Needle Kit, Part Number 9733068 - Passive Biopsy Needle Kit. The image guided Passive Biopsy Needle is an accessory instrument with the StealthStation System. The Biopsy Needle is intended to be used for stereotaxy biopsy of brain tissue.1415 totalClass IITerminated12 Aug 2015
Z-2325-2015Medtronic Passive Biopsy Needle Kit, Part Number 9731754 - Passive Biopsy Fee Per Use (FPU) Kit. The image guided Passive Biopsy Needle is an accessory instrument with the StealthStation System. The Biopsy Needle is intended to be used for stereotaxy biopsy of brain tissue.1415 totalClass IITerminated12 Aug 2015

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2015
1,153
Share of these events in Class II
91%
Product records in this event
2Median of the firm: 1
Events of the firm on the register
44

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 71756. https://recallregister.com/fda/ii/71756/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution - US (nationwide) and Canada.
What quantity of product is in FDA event 71756?
The event has 2 product records. The first record gives this quantity: 1415 total
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2015 have the same class?
91% of the 1,153 device events that started in 2015 are in Class II.