Device recall: MX 16 Slice (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic…
Class II FDA enforcement event 71134. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
The firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Philips Medical Systems (Cleveland) Inc started this device recall on 6 Mar 2015. The firm address in the FDA record is in Cleveland, OH. FDA put the recall in Class II. FDA event 71134 has 35 product records. The first FDA enforcement report date is 17 Jun 2015. The record gives the status Terminated, with the termination date 13 Mar 2017.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide Distribution-US (nationwide) including DC & Puerto Rico and the following countries: Albania, Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia, Brazil, Bulgaria,…
Recalling firm
Philips Medical Systems (Cleveland) Inc, Cleveland, OH.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
35 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1759-2015 | MX 16 Slice (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems. | 125 units | Class II | Terminated | 17 Jun 2015 |
| Z-1760-2015 | MX 16 Slice Phase II (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems. | 11 | Class II | Terminated | 17 Jun 2015 |
| Z-1761-2015 | Brilliance CT 64 channel with Essence, Technology (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems. | 788 units | Class II | Terminated | 17 Jun 2015 |
| Z-1762-2015 | Brilliance CT 40 - Slices (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems. | 17 units | Class II | Terminated | 17 Jun 2015 |
| Z-1763-2015 | Brilliance CT 40 - Slices(Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems. | 376 units | Class II | Terminated | 17 Jun 2015 |
| Z-1764-2015 | Brilliance CT Big Bore Radiology (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems. | 126 units | Class II | Terminated | 17 Jun 2015 |
| Z-1765-2015 | Brilliance CT 16 Slice Air (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems. | 847 units | Class II | Terminated | 17 Jun 2015 |
| Z-1766-2015 | Brilliance CT 6 Slice Air (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems. | 25 units | Class II | Terminated | 17 Jun 2015 |
| Z-1767-2015 | Extended Brilliance Workstation (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems. | 1 unit | Class II | Terminated | 17 Jun 2015 |
| Z-1768-2015 | Brilliance iCT (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems. | 350 units | Class II | Terminated | 17 Jun 2015 |
| Z-1769-2015 | iCT SP (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems. | 110 units | Class II | Terminated | 17 Jun 2015 |
| Z-1770-2015 | Ingenuity Flex (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems. | 23 units | Class II | Terminated | 17 Jun 2015 |
| Z-1771-2015 | Ingenuity Core (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems. | 215 Units | Class II | Terminated | 17 Jun 2015 |
| Z-1772-2015 | Ingenuity Core 128 (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems. | 288 units | Class II | Terminated | 17 Jun 2015 |
| Z-1773-2015 | Ingenuity CT (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems. | 156 units | Class II | Terminated | 17 Jun 2015 |
| Z-1774-2015 | FORTE (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems. | 10 units | Class II | Terminated | 17 Jun 2015 |
| Z-1775-2015 | AXIS (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems. | 1 unit | Class II | Terminated | 17 Jun 2015 |
| Z-1776-2015 | Gemini 16 PET/CT System (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems. | 7 units | Class II | Terminated | 17 Jun 2015 |
| Z-1777-2015 | Forte-JetStream AZ (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems. | 11 units | Class II | Terminated | 17 Jun 2015 |
| Z-1778-2015 | Precedence 16 (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems. | 36 units | Class II | Terminated | 17 Jun 2015 |
| Z-1779-2015 | Precedence 6 (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems. | 1 unit | Class II | Terminated | 17 Jun 2015 |
| Z-1780-2015 | Ingenuity PET/MR (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems. | 9 units | Class II | Terminated | 17 Jun 2015 |
| Z-1781-2015 | Gemini GXL 16 Slice (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems. | 15 units | Class II | Terminated | 17 Jun 2015 |
| Z-1782-2015 | Gemini LXL 16 Slice (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems. | 10 units | Class II | Terminated | 17 Jun 2015 |
| Z-1783-2015 | TruFlight Select PET/CT 16-slice (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems. | 31 units | Class II | Terminated | 17 Jun 2015 |
| Z-1784-2015 | CardioMD III (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems. | 81 units | Class II | Terminated | 17 Jun 2015 |
| Z-1785-2015 | Ingenuity TF (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems. | 14 units | Class II | Terminated | 17 Jun 2015 |
| Z-1786-2015 | GEMINI TF PET/CT 16 Slice (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems. | 69 units | Class II | Terminated | 17 Jun 2015 |
| Z-1787-2015 | GEMINI TF PET/CT 64 Slice (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems. | 71 units | Class II | Terminated | 17 Jun 2015 |
| Z-1788-2015 | GEMINI TF Ready (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems. | 3 units | Class II | Terminated | 17 Jun 2015 |
| Z-1789-2015 | GEMINI TF Big Bore (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems. | 39 units | Class II | Terminated | 17 Jun 2015 |
| Z-1790-2015 | BrightView X (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems. | 108 units | Class II | Terminated | 17 Jun 2015 |
| Z-1791-2015 | BrightView SPECT (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems. | 277 units | Class II | Terminated | 17 Jun 2015 |
| Z-1792-2015 | BrightView XCT (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems. | 337 | Class II | Terminated | 17 Jun 2015 |
| Z-1793-2015 | Diamond Select Brilliance CT 16 (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems. | 2 units | Class II | Terminated | 17 Jun 2015 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2015
- 1,153
- Share of these events in Class II
- 91%
- Product records in this event
- 35Median of the firm: 1
- Events of the firm on the register
- 129
Other events of the firm
Official channels
- FDA enforcement report Z-1759-2015The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 71134. https://recallregister.com/fda/ii/71134/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide Distribution-US (nationwide) including DC & Puerto Rico and the following countries: Albania, Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia, Brazil, Bulgaria,…
- What quantity of product is in FDA event 71134?
- The event has 35 product records. The first record gives this quantity:
125 units
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2015 have the same class?
- 91% of the 1,153 device events that started in 2015 are in Class II.