Device recall: IntelliSpace Portal (ISP) DX/HX/EX, IntelliSpace Portal (ISP) IX, and IntelliSpace Portal (ISP) LX SPECT IntelliSpace…

Class II FDA enforcement event 68953. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

The following software issues have been identified in the affected products. Problem 1: When reopening a bookmark generated from processing a MUGA (Multi-Gated Acquisition) scan within the NM Cardiac Application, the ejection fraction (EF) displayed should be the same as the ejection fraction (EF) originally displayed when the bookmark was first created. In some instances, the ejection fraction (

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Philips Medical Systems (Cleveland) Inc started this device recall on 27 Jul 2014. The firm address in the FDA record is in Cleveland, OH. FDA put the recall in Class II. FDA event 68953 has 1 product record. The first FDA enforcement report date is 20 Aug 2014. The record gives the status Terminated, with the termination date 2 Feb 2016.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution - USA including AK, AL, AR, AZ, CA, CO, FL, GA, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, NC, ND, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, VT, WA, WI, WV, WY,Washington D.C.…

Recalling firm

Philips Medical Systems (Cleveland) Inc, Cleveland, OH.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 68953: product records
Recall numberProductQuantityClassStatusReport date
Z-2237-2014IntelliSpace Portal (ISP) DX/HX/EX, IntelliSpace Portal (ISP) IX, and IntelliSpace Portal (ISP) LX SPECT IntelliSpace Portal (ISP) is a multimodality (CT, NM, and MR) thin-client applications server that delivers full diagnostic viewing and clinical applications to the enterprise. IntelliSpace Portal (ISP) is a…1099 UnitsClass IITerminated20 Aug 2014

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2014
1,176
Share of these events in Class II
91%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
129

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 68953. https://recallregister.com/fda/ii/68953/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution - USA including AK, AL, AR, AZ, CA, CO, FL, GA, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, NC, ND, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, VT, WA, WI, WV, WY,Washington D.C.…
What quantity of product is in FDA event 68953?
The FDA record gives this quantity: 1099 Units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2014 have the same class?
91% of the 1,176 device events that started in 2014 are in Class II.