Device recall: V-Series Patient Monitoring System, Manufactured for Mindray DS USA, Inc. Used for monitoring of human physiological…
Class II FDA enforcement event 68582. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
ECG interpretation feature issues. False measurements reported when the algorithm cannot determine P wave or QRS axis. The Median Complex Report displays the representative beat 25% larger than actual signal input. When a patient date of birth is imported from an external source, the ECG interpretation algorithm uses the default age of 50, and interpretive statements may be incorrect.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Mindray DS USA, Inc. dba Mindray North America started this device recall on 19 May 2014. The firm address in the FDA record is in Mahwah, NJ. FDA put the recall in Class II. FDA event 68582 has 1 product record. The first FDA enforcement report date is 23 Jul 2014. The record gives the status Terminated, with the termination date 20 Apr 2017.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide Distribution - US including the states of Louisiana, New Jersey, Montana, South Carolina and Wyoming.
Recalling firm
Mindray DS USA, Inc. dba Mindray North America, Mahwah, NJ.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
1 product record in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-2059-2014 | V-Series Patient Monitoring System, Manufactured for Mindray DS USA, Inc. Used for monitoring of human physiological parameters. The V-Series Monitor also has the capability of performing IV Drug and Hemodynamic Calculations and interfacing with network devices. | 113 units (57 United States; 56 International) | Class II | Terminated | 23 Jul 2014 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2014
- 1,176
- Share of these events in Class II
- 91%
- Product records in this event
- 1Median of the firm: 1
- Events of the firm on the register
- 29
Other events of the firm
Mindray DS USA, Inc. dba Mindray North America: all 29 events
Official channels
- FDA enforcement report Z-2059-2014The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 68582. https://recallregister.com/fda/ii/68582/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide Distribution - US including the states of Louisiana, New Jersey, Montana, South Carolina and Wyoming.
- What quantity of product is in FDA event 68582?
- The FDA record gives this quantity:
113 units (57 United States; 56 International)
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2014 have the same class?
- 91% of the 1,176 device events that started in 2014 are in Class II.