Food, drugs, devices › Mindray DS USA, Inc. dba Mindray North America
Device recall: Mindray BeneVision Distributed Monitoring System (DMS), Part No. 115-034413-00
Mindray DS USA, Inc. dba Mindray North America (Mahwah, NJ) initiated this device recall on 30 Aug 2018; FDA lists 1 product record under event 81084, classified Class II.
FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 92% of the 1,074 device enforcement events initiated in 2018 carry the same class.
Status: Terminated, terminated 20 Dec 2018. Initiation: voluntary: firm initiated. Mindray DS USA, Inc. dba Mindray North America has 29 enforcement events on this register, with a median of 1 product record per event.
Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.
Recalling firm
Mindray DS USA, Inc. dba Mindray North America, Mahwah, NJ.
Reason for recall
"The firm has identified performance anomalies with the device caused when more than 64 TD60 telemetry transmitters have been connected to a central station. The anomaly causes some ECG waveforms to appear as a flat line at the centralized viewing station and will manifest if more than 64 TD60 transmitters have been connected to a given central station server. Each central station supports only…"
Distribution
"Distributed to accounts in CA, CO, FL, MA, NY, PA, PR, TX, and WV."
Product records
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-0170-2019 | Mindray BeneVision Distributed Monitoring System (DMS), Part No. 115-034413-00 | 5 | Class II | Terminated | 24 Oct 2018 |
Context
Other enforcement events for Mindray DS USA, Inc. dba Mindray North America
All 29 events for Mindray DS USA, Inc. dba Mindray North America
Questions and answers
Why was this device recalled?
FDA's record gives the reason for recall as: "The firm has identified performance anomalies with the device caused when more than 64 TD60 telemetry transmitters have been connected to a central station. The anomaly causes some ECG waveforms to appear as a flat line at the centralized viewing station and will manifest if more than 64 TD60 transmitters have been connected to a given central station server. Each central station supports only…"
What does Class II mean?
FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.
Where was the product distributed?
FDA's distribution pattern field reads: "Distributed to accounts in CA, CO, FL, MA, NY, PA, PR, TX, and WV."
Is the recall still ongoing?
FDA lists the status as Terminated; the termination date is 20 Dec 2018. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.