Device recall: Stryker 0.5mm Round Fluted Bur Sterile Product Usage: The devices are intended to be used to cut bone and bone related…
Class II FDA enforcement event 68331. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
There may be missing notches, partial notches or no notches at all in some Cutting Accessories that are used with the Elite" Attachments.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Stryker Instruments Div. of Stryker Corporation started this device recall on 22 May 2014. The firm address in the FDA record is in Portage, MI. FDA put the recall in Class II. FDA event 68331 has 32 product records. The first FDA enforcement report date is 18 Jun 2014. The record gives the status Terminated, with the termination date 6 May 2015. 4 of the 1,176 device events that started in 2014 have more product records. The median device event of 2014 has 1 product record. The openFDA file has 1,176 device events that started in 2014 and 1,236 that started in 2013.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide Distribution - US (nationwide)
Recalling firm
Stryker Instruments Div. of Stryker Corporation, Portage, MI.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
32 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1749-2014 | Stryker 0.5mm Round Fluted Bur Sterile Product Usage: The devices are intended to be used to cut bone and bone related tissue during neurosurgery, including craniotomy and spinal surgery as well as ear, nose and throat (ENT), orthopedic, and general surgical applications including maxillofacial, craniofacial… | 27 | Class II | Terminated | 18 Jun 2014 |
| Z-1750-2014 | Stryker 1.0mm Round Fluted Bur Sterile Product Usage: The devices are intended to be used to cut bone and bone related tissue during neurosurgery, including craniotomy and spinal surgery as well as ear, nose and throat (ENT), orthopedic, and general surgical applications including maxillofacial, craniofacial… | 72 | Class II | Terminated | 18 Jun 2014 |
| Z-1751-2014 | Stryker 2.0mm Round Fluted Bur Sterile Product Usage: The devices are intended to be used to cut bone and bone related tissue during neurosurgery, including craniotomy and spinal surgery as well as ear, nose and throat (ENT), orthopedic, and general surgical applications including maxillofacial, craniofacial… | 252 | Class II | Terminated | 18 Jun 2014 |
| Z-1752-2014 | Stryker 4.0mm Round Fluted Bur Sterile Product Usage: The devices are intended to be used to cut bone and bone related tissue during neurosurgery, including craniotomy and spinal surgery as well as ear, nose and throat (ENT), orthopedic, and general surgical applications including maxillofacial, craniofacial… | 281 | Class II | Terminated | 18 Jun 2014 |
| Z-1753-2014 | Stryker 6.0mm Round Fluted Bur Sterile Product Usage: The devices are intended to be used to cut bone and bone related tissue during neurosurgery, including craniotomy and spinal surgery as well as ear, nose and throat (ENT), orthopedic, and general surgical applications including maxillofacial, craniofacial… | 256 | Class II | Terminated | 18 Jun 2014 |
| Z-1754-2014 | Stryker 3.0mm Round Fluted Bur Soft Touch Sterile Product Usage: The devices are intended to be used to cut bone and bone related tissue during neurosurgery, including craniotomy and spinal surgery as well as ear, nose and throat (ENT), orthopedic, and general surgical applications including maxillofacial,… | 609 | Class II | Terminated | 18 Jun 2014 |
| Z-1755-2014 | Stryker 5.5mm Round Fluted Bur Soft Touch Sterile Product Usage: The devices are intended to be used to cut bone and bone related tissue during neurosurgery, including craniotomy and spinal surgery as well as ear, nose and throat (ENT), orthopedic, and general surgical applications including maxillofacial,… | 99 | Class II | Terminated | 18 Jun 2014 |
| Z-1756-2014 | Stryker 3.0mm Round Fluted Bur Aggressive Sterile Product Usage: The devices are intended to be used to cut bone and bone related tissue during neurosurgery, including craniotomy and spinal surgery as well as ear, nose and throat (ENT), orthopedic, and general surgical applications including maxillofacial,… | 285 | Class II | Terminated | 18 Jun 2014 |
| Z-1757-2014 | Stryker 4.5mm Round Fluted Bur Aggressive Sterile Product Usage: The devices are intended to be used to cut bone and bone related tissue during neurosurgery, including craniotomy and spinal surgery as well as ear, nose and throat (ENT), orthopedic, and general surgical applications including maxillofacial,… | 62 | Class II | Terminated | 18 Jun 2014 |
| Z-1758-2014 | Stryker 5.0mm Round Fluted Bur Aggressive Sterile Product Usage: The devices are intended to be used to cut bone and bone related tissue during neurosurgery, including craniotomy and spinal surgery as well as ear, nose and throat (ENT), orthopedic, and general surgical applications including maxillofacial,… | 1537 | Class II | Terminated | 18 Jun 2014 |
| Z-1759-2014 | Stryker 5.5mm Round Fluted Bur Aggressive Sterile Product Usage: The devices are intended to be used to cut bone and bone related tissue during neurosurgery, including craniotomy and spinal surgery as well as ear, nose and throat (ENT), orthopedic, and general surgical applications including maxillofacial,… | 84 | Class II | Terminated | 18 Jun 2014 |
| Z-1760-2014 | Stryker 6.0mm Round Fluted Bur Aggressive Sterile Product Usage: The devices are intended to be used to cut bone and bone related tissue during neurosurgery, including craniotomy and spinal surgery as well as ear, nose and throat (ENT), orthopedic, and general surgical applications including maxillofacial,… | 301 | Class II | Terminated | 18 Jun 2014 |
| Z-1761-2014 | Stryker 2.0mm Tapered Round Elite Bur Sterile Product Usage: The devices are intended to be used to cut bone and bone related tissue during neurosurgery, including craniotomy and spinal surgery as well as ear, nose and throat (ENT), orthopedic, and general surgical applications including maxillofacial, craniofacial… | 6 | Class II | Terminated | 18 Jun 2014 |
| Z-1762-2014 | Stryker 3.5mm Zyphr Elite Round Fluted Bur Sterile Product Usage: The devices are intended to be used to cut bone and bone related tissue during neurosurgery, including craniotomy and spinal surgery as well as ear, nose and throat (ENT), orthopedic, and general surgical applications including maxillofacial,… | 39 | Class II | Terminated | 18 Jun 2014 |
| Z-1763-2014 | Stryker 5.0mm Zyphr Elite Round Fluted Bur Sterile Product Usage: The devices are intended to be used to cut bone and bone related tissue during neurosurgery, including craniotomy and spinal surgery as well as ear, nose and throat (ENT), orthopedic, and general surgical applications including maxillofacial,… | 4 | Class II | Terminated | 18 Jun 2014 |
| Z-1764-2014 | Stryker 6.0mm Zyphr Elite Round Fluted Bur Sterile Product Usage: The devices are intended to be used to cut bone and bone related tissue during neurosurgery, including craniotomy and spinal surgery as well as ear, nose and throat (ENT), orthopedic, and general surgical applications including maxillofacial,… | 107 | Class II | Terminated | 18 Jun 2014 |
| Z-1765-2014 | Stryker 2.0mm Zyphr Elite Round Fluted Bur Sterile Product Usage: The devices are intended to be used to cut bone and bone related tissue during neurosurgery, including craniotomy and spinal surgery as well as ear, nose and throat (ENT), orthopedic, and general surgical applications including maxillofacial,… | 4 | Class II | Terminated | 18 Jun 2014 |
| Z-1766-2014 | Stryker 5.0mm Egg Bur Sterile Product Usage: The devices are intended to be used to cut bone and bone related tissue during neurosurgery, including craniotomy and spinal surgery as well as ear, nose and throat (ENT), orthopedic, and general surgical applications including maxillofacial, craniofacial surgeries. The… | 144 | Class II | Terminated | 18 Jun 2014 |
| Z-1767-2014 | Stryker 6.0mm Egg Bur Sterile Product Usage: The devices are intended to be used to cut bone and bone related tissue during neurosurgery, including craniotomy and spinal surgery as well as ear, nose and throat (ENT), orthopedic, and general surgical applications including maxillofacial, craniofacial surgeries. The… | 65 | Class II | Terminated | 18 Jun 2014 |
| Z-1768-2014 | Stryker 5.0mm Barrel Bur Sterile | 15 | Class II | Terminated | 18 Jun 2014 |
| Z-1769-2014 | Stryker 5.0mm Acorn Bur Sterile Product Usage: The devices are intended to be used to cut bone and bone related tissue during neurosurgery, including craniotomy and spinal surgery as well as ear, nose and throat (ENT), orthopedic, and general surgical applications including maxillofacial, craniofacial surgeries. The… | 71 | Class II | Terminated | 18 Jun 2014 |
| Z-1770-2014 | Stryker 6.0mm Acorn Bur Sterile Product Usage: The devices are intended to be used to cut bone and bone related tissue during neurosurgery, including craniotomy and spinal surgery as well as ear, nose and throat (ENT), orthopedic, and general surgical applications including maxillofacial, craniofacial surgeries. The… | 59 | Class II | Terminated | 18 Jun 2014 |
| Z-1771-2014 | Lindemann Drill Short, 1.6mm Sterile Product Usage: The devices are intended to be used to cut bone and bone related tissue during neurosurgery, including craniotomy and spinal surgery as well as ear, nose and throat (ENT), orthopedic, and general surgical applications including maxillofacial, craniofacial surgeries.… | 12 | Class II | Terminated | 18 Jun 2014 |
| Z-1772-2014 | Stryker 1.3mm Wire Pass Drill Sterile Product Usage: The devices are intended to be used to cut bone and bone related tissue during neurosurgery, including craniotomy and spinal surgery as well as ear, nose and throat (ENT), orthopedic, and general surgical applications including maxillofacial, craniofacial… | 176 | Class II | Terminated | 18 Jun 2014 |
| Z-1773-2014 | Stryker 1.5mm Wire Pass Drill Sterile Product Usage: The devices are intended to be used to cut bone and bone related tissue during neurosurgery, including craniotomy and spinal surgery as well as ear, nose and throat (ENT), orthopedic, and general surgical applications including maxillofacial, craniofacial… | 246 | Class II | Terminated | 18 Jun 2014 |
| Z-1774-2014 | Stryker 2.5mm x 3.6mm Neuro Drill Soft Touch Sterile Product Usage: The devices are intended to be used to cut bone and bone related tissue during neurosurgery, including craniotomy and spinal surgery as well as ear, nose and throat (ENT), orthopedic, and general surgical applications including maxillofacial,… | 41 | Class II | Terminated | 18 Jun 2014 |
| Z-1775-2014 | Stryker 3.0mm x 3.8mm Neuro Drill Soft Touch Sterile Product Usage: The devices are intended to be used to cut bone and bone related tissue during neurosurgery, including craniotomy and spinal surgery as well as ear, nose and throat (ENT), orthopedic, and general surgical applications including maxillofacial,… | 802 | Class II | Terminated | 18 Jun 2014 |
| Z-1776-2014 | Stryker 4.0mm x 5.1mm Neuro Drill Soft Touch Sterile Product Usage: The devices are intended to be used to cut bone and bone related tissue during neurosurgery, including craniotomy and spinal surgery as well as ear, nose and throat (ENT), orthopedic, and general surgical applications including maxillofacial,… | 188 | Class II | Terminated | 18 Jun 2014 |
| Z-1777-2014 | Stryker 3.0mm x 3.8mm Neuro Drill Sterile Product Usage: The devices are intended to be used to cut bone and bone related tissue during neurosurgery, including craniotomy and spinal surgery as well as ear, nose and throat (ENT), orthopedic, and general surgical applications including maxillofacial, craniofacial… | 497 | Class II | Terminated | 18 Jun 2014 |
| Z-1778-2014 | Stryker 3.0mm x 3.8mm Precision Neuro Drill Sterile Product Usage: The devices are intended to be used to cut bone and bone related tissue during neurosurgery, including craniotomy and spinal surgery as well as ear, nose and throat (ENT), orthopedic, and general surgical applications including maxillofacial,… | 6481 | Class II | Terminated | 18 Jun 2014 |
| Z-1779-2014 | Stryker 5.0mm Barrel Bur Sterile Product Usage: The devices are intended to be used to cut bone and bone related tissue during neurosurgery, including craniotomy and spinal surgery as well as ear, nose and throat (ENT), orthopedic, and general surgical applications including maxillofacial, craniofacial surgeries. The… | 1 | Class II | Terminated | 18 Jun 2014 |
| Z-1780-2014 | Stryker 3.0mm x 3.8mm Neuro Drill Less Aggressive Sterile Product Usage: The devices are intended to be used to cut bone and bone related tissue during neurosurgery, including craniotomy and spinal surgery as well as ear, nose and throat (ENT), orthopedic, and general surgical applications including maxillofacial,… | 916 | Class II | Terminated | 18 Jun 2014 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2014
- 1,176
- Share of these events in Class II
- 91%
- Product records in this event
- 32Median of the firm: 1
- Events of the firm on the register
- 55
Other events of the firm
Stryker Instruments Div. of Stryker Corporation: all 55 events
Official channels
- FDA enforcement report Z-1749-2014The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 68331. https://recallregister.com/fda/ii/68331/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 10 Nov 2026
- Newest FDA report date on the register
- 4 Nov 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide Distribution - US (nationwide)
- What quantity of product is in FDA event 68331?
- The event has 32 product records. The first record gives this quantity:
27
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2014 have the same class?
- 91% of the 1,176 device events that started in 2014 are in Class II.