Device recall: Stryker TPS Universal Driver REF 5100- 99 Rx , The Stryker Total Performance System (TPS) is intended for use in the…

Class II FDA enforcement event 65456. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

The safety margin values detailed in the Engineering Design were entered incorrectly into the programming software. Handpieces that have been programmed with the incorrect parameters could result in unintended activation. There is a potential for injury to the patient or user as a result of unintended forward or reverse activation. There is also the potential that the handpiece may not reach 100

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Stryker Instruments Div. of Stryker Corporation started this device recall on 1 Jul 2013. The firm address in the FDA record is in Portage, MI. FDA put the recall in Class II. FDA event 65456 has 3 product records. The first FDA enforcement report date is 24 Jul 2013. The record gives the status Terminated, with the termination date 16 Sep 2013.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide distribution: US (nationwide) including states of: AZ, FL, GA, NC, NY, SC, TX and VA; and countries of: India, Netherlands and Japan.

Recalling firm

Stryker Instruments Div. of Stryker Corporation, Portage, MI.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

3 product records in the event.

FDA event 65456: product records
Recall numberProductQuantityClassStatusReport date
Z-1740-2013Stryker TPS Universal Driver REF 5100- 99 Rx , The Stryker Total Performance System (TPS) is intended for use in the cutting, drilling, reaming, decorticating and smoothing of bone and other bone related tissues for a variety of applications such as ENT, dental orthopedic, maxillofacial, spinal, and plastic surgery.…0Class IITerminated24 Jul 2013
Z-1741-2013CORE Universal Driver REF 5400- 99 Rx Only This drill may also be used with theTotal Performance System (TPS").The Stryker Total Performance System (TPS) is intended for usein the cutting, drilling, reaming, decorticating and smoothing of bone and other bone related tissues for a variety of applications such as ENT,…11Class IITerminated24 Jul 2013
Z-1742-2013RemB Electric Universal Driver REF 6400-099-000 For use with the CORE and TPS System sRx Only The RemB Electric Universal Driver is intended for use with the Consolidated Operating Room Equipment (CORE TM) System. When used with a variety of accessories, the RemB Electric Universal Driver is intended for surgical…9Class IITerminated24 Jul 2013

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2013
1,236
Share of these events in Class II
90%
Product records in this event
3Median of the firm: 1
Events of the firm on the register
55

Other events of the firm

Other FDA enforcement events of Stryker Instruments Div. of Stryker Corporation
InitiatedProduct (first record)TypeClassRecords
15 Jun 2022SurgiCount+ System Application, Software Version 2.0.17, Catalog Number 0694-002-090. Used for counting, displaying, recDeviceClass II1
16 Oct 2020Stryker Zyphr Disposable Cranial Perforator, Large 14/11 mm- is a sterile, single use cutting accessory intended for cutDeviceClass II1
26 Mar 2020Neptune E-SEP 165mm Blade Electrode, Catalog Number 0703-165-001 - Product Usage: The Neptune E-SEP is an Integrated SmoDeviceClass II1
22 Jan 2020Triton Canister System (finished part numbers FG 12009, for 3L canisters, and FG 12019 for 2L canisters) Scanning Label DeviceClass II1
13 Jan 2020SafeAir Telescopic Smoke Evacuation Pencil, Product No. 0703 046 004. The product is a Bovie pencil that is 75 inches loDeviceClass II3
6 Nov 2019Color Cuff Disposable Tourniquet Cuff, 18 inch X 3 inch, 1BLA (Single Bladder), 1PRT(Single Port) Quick Connect ConnectoDeviceClass II2
31 Jul 2019C2 Nerve Monitor, 4 Channel, 5140-508-240, UDI 07613327177329 Product Usage: The C2 Nerve Monitor System is intended forDeviceClass II2
19 Mar 2019Performance Series Sagittal Blade, Catalog Number 6125-127-100 Product Usage: The Stryker Heavy Duty Sagittal Blades areDeviceClass II1
4 Dec 2018Triton Sponge System (Software), GAUS-3, Software Versions 4.6.5, 4.8.7, 4.9.6DeviceClass III1
23 Feb 2018T4 Pullover Toga, (S/M); Catalog Number: 0400-710-000 Sterile personal protection garmentDeviceClass II11

Stryker Instruments Div. of Stryker Corporation: all 55 events

Official channels

Cite this page

Recall Register. FDA enforcement event 65456. https://recallregister.com/fda/ii/65456/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide distribution: US (nationwide) including states of: AZ, FL, GA, NC, NY, SC, TX and VA; and countries of: India, Netherlands and Japan.
What quantity of product is in FDA event 65456?
The event has 3 product records. The first record gives this quantity: 0
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2013 have the same class?
90% of the 1,236 device events that started in 2013 are in Class II.