Recall Register

Food, drugs, devices › Alere San Diego, Inc.

Device recall: Alere" hCG Combo Cassette (20/10 mIU/mL) Kit. Product Usage: The AlereTM hCG Combo (20/10 mIU/mL) test is a rapid chroma

FDA enforcement event 65048 · Class II · initiated 16 Apr 2013 · status Terminated · terminated 18 Oct 2013

DeviceVoluntary: Firm initiated1 product recordSan Diego, CA

Alere San Diego, Inc. (San Diego, CA) initiated this device recall on 16 Apr 2013; FDA lists 1 product record under event 65048, classified Class II.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 90% of the 1,236 device enforcement events initiated in 2013 carry the same class.

Status: Terminated, terminated 18 Oct 2013. Initiation: voluntary: firm initiated. Alere San Diego, Inc. has 28 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Alere San Diego, Inc., San Diego, CA.

Reason for recall

"The firm initiated recall of Alere" hCG Combo Cassette (20/10 mIU/mL) kit PN 92215, lots hCG1110133 and hCG1110135, because an investigation has indicated a portion of these Alere" hCG Combo Cassette (20/10 mIU/mL) PN 92215 kit lots were incorrectly packaged with the Alere" hCG Cassette (20 mIU/mL) PN 92210 Product Insert. The Alere" hCG Cassette (20 mIU/mL) Product Insert contains information for"

Distribution

"United States Nationwide Distribution"

Product records

Recall numberProductQuantityClassStatusReport date
Z-1401-2013Alere" hCG Combo Cassette (20/10 mIU/mL) Kit. Product Usage: The AlereTM hCG Combo (20/10 mIU/mL) test is a rapid chromatographic immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in serum or urine, as an aid in the early detection of pregnancy. It is for healthcare professionals only.3975 kitsClass IITerminated5 Jun 2013

Context

90%of 2013 device events are Class II1,236 events that year
1product record in this eventfirm median: 1
28events on this register for the firm

Other enforcement events for Alere San Diego, Inc.

InitiatedProduct (first record)TypeClassRecords
1 Mar 2024Abbott Cholestech LDX Battery Kit, Model: 3LP656194-PCM-LD, PN: 55170DeviceClass II1
24 Jun 2022The Cholestech LDX¿ System combines enzymatic methodology and solidphase technology to measure total cholesterol, HDL chDeviceClass II1
15 Feb 2019Alere Cholestech LDX Analyzer, model # 14-874 containing power supplies with the identifying codes of 1837d and 1843d - DeviceClass II1
14 Nov 2017Triage Drugs of Abuse Plus TCA 25 Test Kit, Model Numbers 92000, 92590, 92591; 25 individually pouched devices, 1 bottleDeviceClass II1
12 Oct 2016Alere Triage Total 5 Control Level 1, Model No. 88753 Description of the product: The Triage Total Controls 5 are qualitDeviceClass II1
2 Aug 2016LDX Optical Check Cassettes (OCC) Model Number: 10-228 and 14-535 Cassette in box with foam insert. The cassette is usedDeviceClass III1
11 Jul 2016Alere INRatio PT/INR System Professional Model Numbers: 0100004 Product Usage: Alere INRatio ¿ 2 PT/INR Monitoring SysteDeviceClass I9
22 Jun 2016Acceava Mono II Rapid Test Strip (Whole Blood, Serum, Plasma), Catalog No. 92404. Mononucleosis Rapid Test Device a simpDeviceClass III7
27 Aug 2015Alere Cholestech LDX Multianalyte Controls, PN 88769 and 88773. Enables users to monitor the performance of total cholesDeviceClass III1
5 Dec 2014Alere INRatio PT/INR Test Strips, Alere INRatio PT/INR System Professional - 0100071 Alere INRatio PT/INR Test Strips, BDeviceClass I3

All 28 events for Alere San Diego, Inc.

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "The firm initiated recall of Alere" hCG Combo Cassette (20/10 mIU/mL) kit PN 92215, lots hCG1110133 and hCG1110135, because an investigation has indicated a portion of these Alere" hCG Combo Cassette (20/10 mIU/mL) PN 92215 kit lots were incorrectly packaged with the Alere" hCG Cassette (20 mIU/mL) PN 92210 Product Insert. The Alere" hCG Cassette (20 mIU/mL) Product Insert contains information for"

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "United States Nationwide Distribution"

Is the recall still ongoing?

FDA lists the status as Terminated; the termination date is 18 Oct 2013. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance