Food, drugs, devices › Alere San Diego, Inc.
Device recall: Abbott Cholestech LDX Battery Kit, Model: 3LP656194-PCM-LD, PN: 55170
Alere San Diego, Inc. (San Diego, CA) initiated this device recall on 1 Mar 2024; FDA lists 1 product record under event 94190, classified Class II.
FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 87% of the 970 device enforcement events initiated in 2024 carry the same class.
Status: Ongoing. Initiation: voluntary: firm initiated. Alere San Diego, Inc. has 28 enforcement events on this register, with a median of 1 product record per event.
Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.
Recalling firm
Alere San Diego, Inc., San Diego, CA.
Reason for recall
"There is a potential that the Cholestech LDX battery (an external battery pack available as an optional accessory to the Cholestech LDX Analyzer) may swell which could cause minor burns, cuts or abrasions, or low current electrical shock to the user."
Distribution
"Worldwide distribution - US Nationwide and the countries of United Kingdom, Netherlands and Kenya."
Product records
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1702-2024 | Abbott Cholestech LDX Battery Kit, Model: 3LP656194-PCM-LD, PN: 55170 | 559 units | Class II | Ongoing | 8 May 2024 |
Context
Other enforcement events for Alere San Diego, Inc.
All 28 events for Alere San Diego, Inc.
Questions and answers
Why was this device recalled?
FDA's record gives the reason for recall as: "There is a potential that the Cholestech LDX battery (an external battery pack available as an optional accessory to the Cholestech LDX Analyzer) may swell which could cause minor burns, cuts or abrasions, or low current electrical shock to the user."
What does Class II mean?
FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.
Where was the product distributed?
FDA's distribution pattern field reads: "Worldwide distribution - US Nationwide and the countries of United Kingdom, Netherlands and Kenya."
Is the recall still ongoing?
FDA lists the status as Ongoing. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.
Where does this data come from?
openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.
Provenance
Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.