Food, drugs, devices › Philips Medical Systems (Cleveland) Inc
Device recall: Brilliance CT 16-Slice Water, System Code: 728245; Brilliance CT 10-Slice Water, System Code: 728250; Brilliance CT 6-Sl
Philips Medical Systems (Cleveland) Inc (Cleveland, OH) initiated this device recall on 12 Mar 2013; FDA lists 1 product record under event 64713, classified Class II.
FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 90% of the 1,236 device enforcement events initiated in 2013 carry the same class.
Status: Terminated, terminated 30 Jun 2015. Initiation: voluntary: firm initiated. Philips Medical Systems (Cleveland) Inc has 129 enforcement events on this register, with a median of 1 product record per event.
Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.
Recalling firm
Philips Medical Systems (Cleveland) Inc, Cleveland, OH.
Reason for recall
"Philips Healthcare received a report from the field stating when they viewed the CTDIVol information from the PACS system, the calculations were too high. Instead of storing the CTDIVol value for each slice in the DICOM tag, the system took the sum of all CTDIVol values from all the images in the dataset and stored this value in the DICOM tag, resulting in incorrect information."
Distribution
"Worldwide distribution: USA (nationwide) including states of: AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK,…"
Product records
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1338-2013 | Brilliance CT 16-Slice Water, System Code: 728245; Brilliance CT 10-Slice Water, System Code: 728250; Brilliance CT 6-Slice Water, System Code: 728255; MX8000 IDT 16, System Code: 728120 & MX8000 IDT 10, System Code: 728125 Computed Tomography X-ray systems. Distributed by Philips Medical Systems (Cleveland), Inc.,… | 515 units | Class II | Terminated | 29 May 2013 |
Context
Other enforcement events for Philips Medical Systems (Cleveland) Inc
All 129 events for Philips Medical Systems (Cleveland) Inc
Questions and answers
Why was this device recalled?
FDA's record gives the reason for recall as: "Philips Healthcare received a report from the field stating when they viewed the CTDIVol information from the PACS system, the calculations were too high. Instead of storing the CTDIVol value for each slice in the DICOM tag, the system took the sum of all CTDIVol values from all the images in the dataset and stored this value in the DICOM tag, resulting in incorrect information."
What does Class II mean?
FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.
Where was the product distributed?
FDA's distribution pattern field reads: "Worldwide distribution: USA (nationwide) including states of: AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK,…"
Is the recall still ongoing?
FDA lists the status as Terminated; the termination date is 30 Jun 2015. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.