Device recall: Integra CUSA excel. CEM Nosecone for CUSA EXcel System. The CEM Nosecone is intended for use CUSA Excel/CUSA Excel +…

Class II FDA enforcement event 64318. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

There is a potential for erosion of the CUSA EXcel tip used with the Nosecone. Specifically, the potential for tip erosion occurs when using a CUSA CEM Nosecone if the side of the tip, rather than the annulus of the tip, is in contact with the tissue. This erosion can potentially lead to tip breakage. There were no reports of a patient injury associated with these complaints.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Integra LifeSciences Corp. started this device recall on 7 Jan 2013. The firm address in the FDA record is in Plainsboro, NJ. FDA put the recall in Class II. FDA event 64318 has 1 product record. The first FDA enforcement report date is 20 Mar 2013. The record gives the status Terminated, with the termination date 6 Aug 2013.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution - USA and the countries of Austria, Australia, Bangladesh, Belgium, Bulgaria, Brazil, Switzerland, Canada, China, Costa Rica, Czech Republic, Germany, Dominican Republic,…

Recalling firm

Integra LifeSciences Corp., Plainsboro, NJ.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 64318: product records
Recall numberProductQuantityClassStatusReport date
Z-0920-2013Integra CUSA excel. CEM Nosecone for CUSA EXcel System. The CEM Nosecone is intended for use CUSA Excel/CUSA Excel + Ultrasonic Surgical Aspiration System, CUSA handpiece, and Covidien Force FX Electrosurgical Generator. Used in surgical procedures where combined ultrasonic dissection and electrosurgical coagulation…USA 6,190, EU 22,972 and ROW 288 unitsClass IITerminated20 Mar 2013

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2013
1,236
Share of these events in Class II
90%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
79

Other events of the firm

Other FDA enforcement events of Integra LifeSciences Corp.
InitiatedProduct (first record)TypeClassRecords
6 Feb 2025AURORA Surgiscope System, Sterile, single use device that contains a Sheath, Obturator, and Imager. Manufacturer's ProduDeviceClass II1
10 Jan 2025Brand Name: MediHoney Gel Product Name: MediHoney Wound & Burn Dressing Gel Model/Catalog Number: 31815 Software VersionDeviceClass II1
16 Dec 2024CODMAN CERTAS Plus Programmable Valve Inline Small Valve Only. Model Number: 828810 Implantable device that provides conDeviceClass II3
26 Sep 2024Duraform Dural Graft Implant, Size 3x3 , Model Number 80-1477US Lot CT007694DeviceClass II1
20 Sep 2024MICR PATIE RND Model/Catalog Number: 801396. CODMAN Surgical Patties and CODMAN Surgical Strips are manufactured of COTTDeviceClass II22
13 Sep 2024MEDIHONEY Dressing with Active Leptospermum Honey Wound and Burn Dressing W/APPLICATOR, 1.5oz tube (Model Number 31515) DeviceClass II1
9 Jul 2024TruDi NAV Suction, 0 Degrees; Model/Catalog Number: TDNS000Z. TruDi" NAV Suction Instruments are intended for use with tDeviceClass II1
7 Jun 2024AURORA Surgiscope System. Sterile, single use device that contains a Sheath, Obturator, and Imager.DeviceClass II1
29 May 2024CUSA Excel C2600 23KHz Straight Handpiece-Indicated for use in surgical procedures where fragmentation, emulsification aDeviceClass II1
7 May 2024CODMAN BACTISEAL EVD Catheter Set - Indicated for gaining access to the ventricles of the brain and can be used with dimDeviceClass II3

Integra LifeSciences Corp.: all 79 events

Official channels

Cite this page

Recall Register. FDA enforcement event 64318. https://recallregister.com/fda/ii/64318/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution - USA and the countries of Austria, Australia, Bangladesh, Belgium, Bulgaria, Brazil, Switzerland, Canada, China, Costa Rica, Czech Republic, Germany, Dominican Republic,…
What quantity of product is in FDA event 64318?
The FDA record gives this quantity: USA 6,190, EU 22,972 and ROW 288 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2013 have the same class?
90% of the 1,236 device events that started in 2013 are in Class II.