Device recall: MICR PATIE RND Model/Catalog Number: 801396. CODMAN Surgical Patties and CODMAN Surgical Strips are manufactured of…
Class II FDA enforcement event 95432. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Integra LifeSciences Corp. started this device recall on 20 Sep 2024. The firm address in the FDA record is in Princeton, NJ. FDA put the recall in Class II. FDA event 95432 has 22 product records. The first FDA enforcement report date is 6 Nov 2024. The record gives the status Ongoing. 25 of the 972 device events that started in 2024 have more product records. The median device event of 2024 has 1 product record. The openFDA file has 972 device events that started in 2024 and 917 that started in 2023.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide Distribution. US Nationwide, Canada Germany United Kingdom of Great Britain and Northern Ireland (the) Malta Italy Barbados Trinidad and Tobago Australia New Zealand Netherlands (the)…
Recalling firm
Integra LifeSciences Corp., Princeton, NJ.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
22 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-0235-2025 | MICR PATIE RND Model/Catalog Number: 801396. CODMAN Surgical Patties and CODMAN Surgical Strips are manufactured of COTTONOID Material with x-ray detectable markers. All patties have a suture string attached for ease in performing postsurgical count verification. The surgical patties and surgical strips are… | 1791 | Class II | Ongoing | 6 Nov 2024 |
| Z-0236-2025 | SURG PAT XRAY 1/4X1-1/2 Model/Catalog Number: 801397. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery. | 2615 | Class II | Ongoing | 6 Nov 2024 |
| Z-0237-2025 | SURG PAT XRAY 1/4X3 Model/Catalog Number: 801398. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery. | 2894 | Class II | Ongoing | 6 Nov 2024 |
| Z-0238-2025 | SURG PAT XRAY 1/4X1/4 Model/Catalog Number: 801399. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery. | 21553 | Class II | Ongoing | 6 Nov 2024 |
| Z-0239-2025 | SURG PAT XRAY 1/2X1/2 Model/Catalog Number: 801400. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery. | 168803 | Class II | Ongoing | 6 Nov 2024 |
| Z-0240-2025 | SURG PAT XRAY 3/4X3/4 Model/Catalog Number: 801401. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery. | 33478 | Class II | Ongoing | 6 Nov 2024 |
| Z-0241-2025 | SURG PAT XRAY 1/2X1 Model/Catalog Number: 801402. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery. | 80515 | Class II | Ongoing | 6 Nov 2024 |
| Z-0242-2025 | SURG PAT XRAY 1X1 Model/Catalog Number: 801403. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery. | 44231 | Class II | Ongoing | 6 Nov 2024 |
| Z-0243-2025 | SURG PAT XRAY 1/2X1-1/2 Model/Catalog Number: 801404. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery. | 69179 | Class II | Ongoing | 6 Nov 2024 |
| Z-0244-2025 | SURG PAT XRAY 1/2X2 Model/Catalog Number: 801406. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery. | 56082 | Class II | Ongoing | 6 Nov 2024 |
| Z-0245-2025 | SURG PAT XRAY 1/2X3 Model/Catalog Number: 801407. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery. | 201914 | Class II | Ongoing | 6 Nov 2024 |
| Z-0246-2025 | SURG PAT XRAY 1X3 Model/Catalog Number: 801408. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery. | 93782 | Class II | Ongoing | 6 Nov 2024 |
| Z-0247-2025 | SURG PAT XRAY 3X3 Model/Catalog Number: 801409. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery. | 9563 | Class II | Ongoing | 6 Nov 2024 |
| Z-0248-2025 | SURG STRP 1/8X6 Model/Catalog Number: 801449. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery. | 826 | Class II | Ongoing | 6 Nov 2024 |
| Z-0249-2025 | SURG STRP 1/4X6 Model/Catalog Number: 801450. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery. | 6706 | Class II | Ongoing | 6 Nov 2024 |
| Z-0250-2025 | SURG STRP 1/2X6 Model/Catalog Number: 801451. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery. | N/A | Class II | Ongoing | 6 Nov 2024 |
| Z-0251-2025 | SURG STRP 3/4X6 Model/Catalog Number: 801452. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery. | 4818 | Class II | Ongoing | 6 Nov 2024 |
| Z-0252-2025 | SURG STRP 1X6-200 Model/Catalog Number: 801453. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery. | 10719 | Class II | Ongoing | 6 Nov 2024 |
| Z-0253-2025 | SURG STRP 1-1/2X6 Model/Catalog Number: 801454. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery. | 2056 | Class II | Ongoing | 6 Nov 2024 |
| Z-0254-2025 | SURG STRP 2X6 Model/Catalog Number: 801455. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery. | 1222 | Class II | Ongoing | 6 Nov 2024 |
| Z-0255-2025 | SURG STRP 3X6 Model/Catalog Number: 801456. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery. | 1454 units | Class II | Ongoing | 6 Nov 2024 |
| Z-0256-2025 | SURG STRP 3-1/2X6 Model/Catalog Number:801457. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery. | 3494 units | Class II | Ongoing | 6 Nov 2024 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2024
- 972
- Share of these events in Class II
- 87%
- Product records in this event
- 22Median of the firm: 1
- Events of the firm on the register
- 79
Other events of the firm
Official channels
- FDA enforcement report Z-0235-2025The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 95432. https://recallregister.com/fda/ii/95432/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 10 Nov 2026
- Newest FDA report date on the register
- 4 Nov 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide Distribution. US Nationwide, Canada Germany United Kingdom of Great Britain and Northern Ireland (the) Malta Italy Barbados Trinidad and Tobago Australia New Zealand Netherlands (the)…
- What quantity of product is in FDA event 95432?
- The event has 22 product records. The first record gives this quantity:
1791
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2024 have the same class?
- 87% of the 972 device events that started in 2024 are in Class II.