Recall Register

Device recall: Ortho Clinical Diagnostics VITROS Chemistry Products Vapor Adsorption Cartridge REF/CAT No. 680 0100. VITROS Vapor…

Class II FDA enforcement event 63934. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Some VITROS Chemistry Products Vapor Adsorption Cartridge CAT No. 6800100 cartons incorrectly contain a Particulate Cartridge Part No. J19612. In some instances, the Particulate Cartridge has been inadvertently installed on the VITROS System instead of the intended VITROS Vapor Adsorption Cartridge which may bias results.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Ortho-Clinical Diagnostics started this device recall on 13 Dec 2012. The firm address in the FDA record is in Rochester, NY. FDA put the recall in Class II. FDA event 63934 has 1 product record. The first FDA enforcement report date is 6 Feb 2013. The record gives the status Terminated, with the termination date 5 Jul 2018. 182 of the 823 device events that started in 2012 have more product records. The median device event of 2012 has 1 product record. The openFDA file has 823 device events that started in 2012 and 114 that started in 2011.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution, including Nationwide (USA) and the countries of AU, GM, FR, IT, SP, BR, BM, AR, CA, JA, MX, UK, CH, CL, CO, IN, EC, PE, PY, PA, PT, NZ, SG, VE and UY.

Recalling firm

Ortho-Clinical Diagnostics, Rochester, NY.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 63934: product records
Recall numberProductQuantityClassStatusReport date
Z-0755-2013Ortho Clinical Diagnostics VITROS Chemistry Products Vapor Adsorption Cartridge REF/CAT No. 680 0100. VITROS Vapor Adsorption Cartridges are used to remove vapor contaminants that may potentially interfere with Signal Reagent used with MicroWell Assays on the VITROS 3600, VITROS 5600, and VITROS ECi/ECiQ Systems.Domestic: 15,891; Foreign: 5237Class IITerminated6 Feb 2013

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2012
823
Share of these events in Class II
88%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
105

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 63934. https://recallregister.com/fda/ii/63934/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
10 Nov 2026
Newest FDA report date on the register
4 Nov 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution, including Nationwide (USA) and the countries of AU, GM, FR, IT, SP, BR, BM, AR, CA, JA, MX, UK, CH, CL, CO, IN, EC, PE, PY, PA, PT, NZ, SG, VE and UY.
What quantity of product is in FDA event 63934?
The FDA record gives this quantity: Domestic: 15,891; Foreign: 5237
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2012 have the same class?
88% of the 823 device events that started in 2012 are in Class II.