Recall Register

Device recall: Alere Triage TOX Drug Screen 9 Panel, P/N: 94402. The Alere Triage TOX Drug Screen is a fluorescence immunoassay…

Class II FDA enforcement event 63091. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Alere San Diego is recalling Alere Triage TOX Drug Screen because there is a possibility that after installing the code chip provided with the affected lots: AMP, BAR, and PCP, the threshold ranges may not be consistently set correctly in all meters.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Alere San Diego, Inc. started this device recall on 12 Jul 2012. The firm address in the FDA record is in San Diego, CA. FDA put the recall in Class II. FDA event 63091 has 1 product record. The first FDA enforcement report date is 3 Oct 2012. The record gives the status Terminated, with the termination date 1 Feb 2013. 182 of the 823 device events that started in 2012 have more product records. The median device event of 2012 has 1 product record. The openFDA file has 823 device events that started in 2012 and 114 that started in 2011.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Nationwide Distribution.

Recalling firm

Alere San Diego, Inc., San Diego, CA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 63091: product records
Recall numberProductQuantityClassStatusReport date
Z-2460-2012Alere Triage TOX Drug Screen 9 Panel, P/N: 94402. The Alere Triage TOX Drug Screen is a fluorescence immunoassay intended to be used with the Alere Triage Meters for the qualitative determination of the presence of drug and/or the major metabolites above the threshold concentrations of up to 10 distinct drug classes,…591 kitsClass IITerminated3 Oct 2012

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2012
823
Share of these events in Class II
88%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
29

Other events of the firm

Other FDA enforcement events of Alere San Diego, Inc.
InitiatedProduct (first record)TypeClassRecords
20 Aug 2026McKesson CONSULT hCG Urine Cassette, Model/ Part: FHC-A102/5001; Alere hCG Cassette (20 mIU/mL) in Urine, Model/ Part: FDeviceClass II4
1 Mar 2024Abbott Cholestech LDX Battery Kit, Model: 3LP656194-PCM-LD, PN: 55170DeviceClass II1
24 Jun 2022The Cholestech LDX System combines enzymatic methodology and solidphase technology to measure total cholesterol, HDL choDeviceClass II1
15 Feb 2019Alere Cholestech LDX Analyzer, model # 14-874 containing power supplies with the identifying codes of 1837d and 1843d - DeviceClass II1
14 Nov 2017Triage Drugs of Abuse Plus TCA 25 Test Kit, Model Numbers 92000, 92590, 92591; 25 individually pouched devices, 1 bottleDeviceClass II1
12 Oct 2016Alere Triage Total 5 Control Level 1, Model No. 88753 Description of the product: The Triage Total Controls 5 are qualitDeviceClass II1
2 Aug 2016LDX Optical Check Cassettes (OCC) Model Number: 10-228 and 14-535 Cassette in box with foam insert. The cassette is usedDeviceClass III1
11 Jul 2016Alere INRatio PT/INR System Professional Model Numbers: 0100004 Product Usage: Alere INRatio 2 PT/INR Monitoring System DeviceClass I9
22 Jun 2016Acceava Mono II Rapid Test Strip (Whole Blood, Serum, Plasma), Catalog No. 92404. Mononucleosis Rapid Test Device a simpDeviceClass III7
27 Aug 2015Alere Cholestech LDX Multianalyte Controls, PN 88769 and 88773. Enables users to monitor the performance of total cholesDeviceClass III1

Alere San Diego, Inc.: all 29 events

Official channels

Cite this page

Recall Register. FDA enforcement event 63091. https://recallregister.com/fda/ii/63091/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
10 Nov 2026
Newest FDA report date on the register
4 Nov 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Nationwide Distribution.
What quantity of product is in FDA event 63091?
The FDA record gives this quantity: 591 kits
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2012 have the same class?
88% of the 823 device events that started in 2012 are in Class II.