Device recall: Alere Triage TOX Drug Screen 9 Panel, P/N: 94402. The Alere Triage TOX Drug Screen is a fluorescence immunoassay intende
Alere San Diego, Inc. (San Diego, CA) initiated this device recall on 12 Jul 2012; FDA lists 1 product record under event 63091, classified Class II.
FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 88% of the 823 device enforcement events initiated in 2012 carry the same class.
Status: Terminated, terminated 1 Feb 2013. Initiation: voluntary: firm initiated. Alere San Diego, Inc. has 28 enforcement events on this register, with a median of 1 product record per event.
Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.
Recalling firm
Alere San Diego, Inc., San Diego, CA.
Reason for recall
"Alere San Diego is recalling Alere Triage TOX Drug Screen because there is a possibility that after installing the code chip provided with the affected lots: AMP, BAR, and PCP, the threshold ranges may not be consistently set correctly in all meters."
Distribution
"Nationwide Distribution."
Product records
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-2460-2012 | Alere Triage TOX Drug Screen 9 Panel, P/N: 94402. The Alere Triage TOX Drug Screen is a fluorescence immunoassay intended to be used with the Alere Triage Meters for the qualitative determination of the presence of drug and/or the major metabolites above the threshold concentrations of up to 10 distinct drug classes,… | 591 kits | Class II | Terminated | 3 Oct 2012 |
Context
Other enforcement events for Alere San Diego, Inc.
All 28 events for Alere San Diego, Inc.
Questions and answers
Why was this device recalled?
FDA's record gives the reason for recall as: "Alere San Diego is recalling Alere Triage TOX Drug Screen because there is a possibility that after installing the code chip provided with the affected lots: AMP, BAR, and PCP, the threshold ranges may not be consistently set correctly in all meters."
What does Class II mean?
FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.
Where was the product distributed?
FDA's distribution pattern field reads: "Nationwide Distribution."
Is the recall still ongoing?
FDA lists the status as Terminated; the termination date is 1 Feb 2013. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.