Recall Register

Device recall: OPTETRAK LOGIC TIBIAL INSERT POSTERIOR STABILIZED***Match Component Colors 6 YELLOW, 11mm, Use with LOGIC Femoral and…

Class II FDA enforcement event 61984. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Exactech, Inc. of Gainesville, FL is recalling their Optetrak Logic Posterior Stabilized Tibial Insert, Size 6 after the device was determined to be out of dimensional specification.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Exactech, Inc. started this device recall on 19 Aug 2011. The firm address in the FDA record is in Gainesville, FL. FDA put the recall in Class II. FDA event 61984 has 1 product record. The first FDA enforcement report date is 12 Sep 2012. The record gives the status Terminated, with the termination date 7 Nov 2013.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

USA ( nationwide) WA, OH, OK, TX, NC, NY, CO, VA, MA, IL, TN, MN, NJ and FL.

Recalling firm

Exactech, Inc., Gainesville, FL.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 61984: product records
Recall numberProductQuantityClassStatusReport date
Z-2310-2012OPTETRAK LOGIC TIBIAL INSERT POSTERIOR STABILIZED***Match Component Colors 6 YELLOW, 11mm, Use with LOGIC Femoral and Tibial Components***REF 02-012-35-6011. There are three product sizes being recalled, 11mm, 13mm, 15mm. All other information is the same. The Optetrak Logic Total Knee System is indicated for use in…193Class IITerminated12 Sep 2012

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2011
114
Share of these events in Class II
89%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
61

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 61984. https://recallregister.com/fda/ii/61984/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: USA ( nationwide) WA, OH, OK, TX, NC, NY, CO, VA, MA, IL, TN, MN, NJ and FL.
What quantity of product is in FDA event 61984?
The FDA record gives this quantity: 193
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2011 have the same class?
89% of the 114 device events that started in 2011 are in Class II.