Device recall: Exactech Equinoxe GLENOID,KEELED, CEMENTED, Mates with 38, 41, 44 head, alpha curvature, Sizes: a), Small, Item Number…
Class II FDA enforcement event 94092. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Exactech, Inc. started this device recall on 6 Mar 2024. The firm address in the FDA record is in Gainesville, FL. FDA put the recall in Class II. FDA event 94092 has 24 product records. The first FDA enforcement report date is 10 Apr 2024. The record gives the status Ongoing.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Nationwide and Australia, Austria, Belgium, Brazil, Chile, Canada , China, Colombia, France, Germany, Great Britain, Greece, Guatemala, India, Italy, Japan, Korea, Luxembourg, Malaysia, Netherlands,…
Recalling firm
Exactech, Inc., Gainesville, FL.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
24 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1402-2024 | Exactech Equinoxe GLENOID,KEELED, CEMENTED, Mates with 38, 41, 44 head, alpha curvature, Sizes: a), Small, Item Number 314-01-02; b) Medium, Item number 314-01-03; c) Large, Item Number 314-01-04, Shoulder Arthroplasty | 2077 units | Class II | Ongoing | 10 Apr 2024 |
| Z-1403-2024 | Exactech Equinoxe GLENOID,KEELED, CEMENTED, mates with 47, 50, 53 head, beta curvature, Sizes: a) Small, Item Number 314-01-12, b) Medium, Item Number 314-01-13, c) Large, Item Number 314-01-14, Shoulder Arthroplasty | 2008 units | Class II | Ongoing | 10 Apr 2024 |
| Z-1404-2024 | Exactech Equinoxe GLENOID,KEELED, CEMENTED, mates with 47, 50, 53 head, beta curvature, Sizes: a) Small, Item Number 314-02-02, b) Medium, Item Number 314-02-03, c) Large, Item Number 314-02-04, Shoulder Arthroplasty | 6678 units | Class II | Ongoing | 10 Apr 2024 |
| Z-1405-2024 | Exactech Equinoxe GLENOID,KEELED, CEMENTED,47, 50, 53 head, beta curvature, Sizes: a) Small, Item Number 314-02-12, b) Medium, Item Number 314-02-13, c) Large, Item Number 314-02-14, d) Extra Large, Item Number 314-02-15, Shoulder Arthroplasty | 6840 units | Class II | Ongoing | 10 Apr 2024 |
| Z-1406-2024 | Exactech Equinoxe GLENOID, POSTERIOR AUGMENT,PEGGED, CEMENTED, 8o, LEFT, Mates with: a) 38, 41, 44, 47 head, Small, Item Number 314-02-22, b) 41, 44, 47, 50 head, Medium, Item Number 314-02-23, c) 44, 47, 50, 53 head, Large, Item Number 314-02-24, d) 44, 47, 50, 53 head, Extra Large, Item Number 314-02-25; Shoulder… | 805 units | Class II | Ongoing | 10 Apr 2024 |
| Z-1407-2024 | Exactech Equinoxe GLENOID, POSTERIOR AUGMENT,PEGGED, CEMENTED, 8o, RIGHT, Mates with: a) 38, 41, 44, 47 head, Small, Item Number 314-02-32, b) 41, 44, 47, 50 head, Medium, Item Number 314-02-33, c) 44, 47, 50, 53 head, Large, Item Number 314-02-34, d) 44, 47, 50, 53 head, Extra Large, Item Number 314-02-35; Shoulder… | 985 units | Class II | Ongoing | 10 Apr 2024 |
| Z-1408-2024 | Exactech Equinoxe GLENOID,POSTERIOR AUGMENT, PEGGED, 12o, LEFT, FOR CEMENTED USE ONLY, Mates with: a) 38, 41, 44, 47 head, Small, Item Number 314-04-22, b) 41, 44, 47, 50 head, Medium, Item Number 314-04-23, c) 44, 47, 50, 53 head, Large, Item Number 314-04-24, d) 44, 47, 50, 53 head, Extra Large, Item Number… | 35 units | Class II | Ongoing | 10 Apr 2024 |
| Z-1409-2024 | Exactech Equinoxe GLENOID,POSTERIOR AUGMENT, PEGGED, 12o, RIGHT, FOR CEMENTED USE ONLY, Mates with: a) 38, 41, 44, 47 head, Small, Item Number 314-04-32, b) 41, 44, 47, 50 head, Medium, Item Number 314-04-33, c) 44, 47, 50, 53 head, Large, Item Number 314-04-34, d) 44, 47, 50, 53 head, Extra Large, Item Number… | 39 units | Class II | Ongoing | 10 Apr 2024 |
| Z-1410-2024 | Exactech Equinoxe GLENOID,POSTERIOR AUGMENT, PEGGED, 16o, LEFT, FOR CEMENTED USE ONLY, Mates with: a) 38, 41, 44, 47 head, Small, Item Number 314-06-22, b) 41, 44, 47, 50 head, Medium, Item Number 314-06-23, c) 44, 47, 50, 53 head, Large, Item Number 314-06-24, d) 44, 47, 50, 53 head, Extra Large, Item Number… | 180 units | Class II | Ongoing | 10 Apr 2024 |
| Z-1411-2024 | Exactech Equinoxe GLENOID,POSTERIOR AUGMENT, PEGGED, 16o, RIGHT, FOR CEMENTED USE ONLY, Mates with: a) 38, 41, 44, 47 head, Small, Item Number 314-06-32, b) 41, 44, 47, 50 head, Medium, Item Number314-06-33, c) 44, 47, 50, 53 head, Large, Item Number 314-06-34, d) 44, 47, 50, 53 head, Extra Large, Item Number… | 245 units | Class II | Ongoing | 10 Apr 2024 |
| Z-1412-2024 | Exactech Equinoxe CAGE GLENOID, CEMENTED, alpha curvature, Mates with: a) 38, 41, 44 head, Small, Item Number 314-13-02, b) 38, 41, 44 head, Medium, Item Number 314-13-03, c) 38, 41, 44 head, Large, Item Number 314-13-04; Shoulder Arthroplasty | 11031 units | Class II | Ongoing | 10 Apr 2024 |
| Z-1413-2024 | Exactech Equinoxe CAGE GLENOID, CEMENTED, beta curvature, Mates with: a) 47, 50, 53 head, Medium, Item Number 314-13-13, b) 47, 50, 53 head, Large, Item Number 314-13-14, c) 47, 50, 53 head, Extra Large, Item Number 314-13-15; Shoulder Arthroplasty | 12643 units | Class II | Ongoing | 10 Apr 2024 |
| Z-1414-2024 | Exactech Equinoxe CAGE GLENOID,POSTERIOR AUGMENT, CEMENTED, Left, Mates with: a) 38, 41, 44, 47 head, Small, Item Number 314-13-22, b) 41, 44, 47, 50, head, Medium, Item Number 314-13-23, c) 44, 47, 50, 53 head, Large, Item Number 314-13-24, d) 44, 47, 50, 53 head, Extra Large, Item Number 314-13-25; Shoulder… | 3436 units | Class II | Ongoing | 10 Apr 2024 |
| Z-1415-2024 | Exactech Equinoxe CAGE GLENOID,POSTERIOR AUGMENT, CEMENTED, Right, Mates with: a) 38, 41, 44, 47 head, Small, Item Number 314-13-32, b) 41, 44, 47, 50, head, Medium, Item Number 314-13-33, c) 44, 47, 50, 53 head, Large, Item Number 314-13-34, d) 44, 47, 50, 53 head, Extra Large, Item Number 314-13-35; Shoulder… | 4007 units | Class II | Ongoing | 10 Apr 2024 |
| Z-1416-2024 | Exactech Equinoxe REVERSE SHOULDER,36mm Humeral Liner: a) +0mm, Item Number 320-36-00, b) +2.5mm, Item Number 320-36-03; Shoulder Arthroplasty | 11863 units | Class II | Ongoing | 10 Apr 2024 |
| Z-1417-2024 | Exactech Equinoxe REVERSE SHOULDER,36mm Constrained Humeral Liner: a) +0mm, Item Number 320-36-10, b) +2.5mm, Item Number 320-36-13; Shoulder Arthroplasty | 826 units | Class II | Ongoing | 10 Apr 2024 |
| Z-1418-2024 | Exactech Equinoxe REVERSE SHOULDER, 38mm Humeral Liner: a) +0mm, Item Number 320-38-00, b) +2.5mm, Item Number 320-38-03; Shoulder Arthroplasty | 66646 units | Class II | Ongoing | 10 Apr 2024 |
| Z-1419-2024 | Exactech Equinoxe REVERSE SHOULDER,38mm Constrained Humeral Liner: a) +0mm, Item Number 320-38-10, b) +2.5mm, Item Number 320-38-13, Shoulder Arthroplasty | 7213 units | Class II | Ongoing | 10 Apr 2024 |
| Z-1420-2024 | Exactech Equinoxe Humeral Liner: Item Numbers: a) 320-40-00, b) 320-40-03; Shoulder Arthroplasty | 2806 units | Class II | Ongoing | 10 Apr 2024 |
| Z-1421-2024 | Exactech Equinoxe Constrained Humeral Liner: Item Numbers: a) 320-40-10, b) 320-40-13, Shoulder Arthroplasty | 447 units | Class II | Ongoing | 10 Apr 2024 |
| Z-1422-2024 | Exactech Equinoxe REVERSE SHOULDER,42mm Humeral Liner, a) +0mm, Item number 320-42-00, b) +2.5mm, Item Number 320-42-03; Shoulder Arthroplasty | 24303 units | Class II | Ongoing | 10 Apr 2024 |
| Z-1423-2024 | Exactech Equinoxe REVERSE SHOULDER,42mm Constrained Humeral Liner, a) +0mm, Item number 320-42-10, b) +2.5mm, Item Number 320-42-13; Shoulder Arthroplasty | 3648 units | Class II | Ongoing | 10 Apr 2024 |
| Z-1424-2024 | Exactech Equinoxe REVERSE SHOULDER,46mm Humeral Liner, a) +0mm, Item number 320-46-00, b) +2.5mm, Item Number 320-46-03; Shoulder Arthroplasty | 1970 units | Class II | Ongoing | 10 Apr 2024 |
| Z-1425-2024 | Exactech Equinoxe REVERSE SHOULDER,46mmConstrained Humeral Liner, a) +0mm, Item number 320-46-10, b) +2.5mm, Item Number 320-46-13; Shoulder Arthroplasty | 591 units | Class II | Ongoing | 10 Apr 2024 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2024
- 970
- Share of these events in Class II
- 87%
- Product records in this event
- 24Median of the firm: 1
- Events of the firm on the register
- 61
Other events of the firm
Official channels
- FDA enforcement report Z-1402-2024The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 94092. https://recallregister.com/fda/ii/94092/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Nationwide and Australia, Austria, Belgium, Brazil, Chile, Canada , China, Colombia, France, Germany, Great Britain, Greece, Guatemala, India, Italy, Japan, Korea, Luxembourg, Malaysia, Netherlands,…
- What quantity of product is in FDA event 94092?
- The event has 24 product records. The first record gives this quantity:
2077 units
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2024 have the same class?
- 87% of the 970 device events that started in 2024 are in Class II.