Device recall: Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM Android App Model/Catalog…
Class I FDA enforcement event 97262. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
The affected devices are the Dexcom G7 Continuous Glucose Monitoring System (CGM System) using iOS, watchOS and Android App software (versions 2.8.0 or earlier) and Dexcom ONE+ CGM System using iOS and Android App software (versions 1.4.0 or earlier), henceforth referred to as the “App”. The affected App does not provide an expected "Sensor Failed" alert when the transmitter sends a…
Class I, in the words of FDA
a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death
Dexcom, Inc. started this device recall on 24 Jul 2025. The firm address in the FDA record is in San Diego, CA. FDA put the recall in Class I. FDA event 97262 has 5 product records. The first FDA enforcement report date is 10 Sep 2025. The record gives the status Ongoing.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
SW12299 G7 Android CGM App Worldwide Distribution - US Nationwide and the countries of Andorra, United Arab Emirates, Austria, Australia, Belgium, Bahrain, Canada, Switzerland, Czech Republic,…
Recalling firm
Dexcom, Inc., San Diego, CA.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
5 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-2446-2025 | Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM Android App Model/Catalog Number: SW12299 Software Version: version 2.8.0 and earlier Product Description: Dexcom G7 CGM Android App SW12299 Component: Dexcom G7 Continuous Glucose Monitoring System | 532821 | Class I | Ongoing | 10 Sep 2025 |
| Z-2447-2025 | Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM iOS App Model/Catalog Number: SW12300 Software Version: version 2.8.0 and earlier Product Description: Dexcom G7 CGM iOS App Component: Dexcom G7 Continuous Glucose Monitoring System | 907842 | Class I | Ongoing | 10 Sep 2025 |
| Z-2448-2025 | Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM watchOS App Model/Catalog Number: SW13355 Software Version: versions 2.8.0 and earlier Product Description: Dexcom G7 CGM watchOS App - The watchOS app is installed along with the iOS phone app and the versions used of each are the… | 907842 | Class I | Ongoing | 10 Sep 2025 |
| Z-2449-2025 | Brand Name: Dexcom ONE+ Continuous Glucose Monitoring System Product Name: Dexcom ONE+ Android CGM App Model/Catalog Number: SW14245 Software Version: versions 1.4.0 and earlier Product Description: Dexcom ONE+ Androind CGM App Component: Dexcom ONE+ Continuous Glucose Monitoring System | 41629 | Class I | Ongoing | 10 Sep 2025 |
| Z-2450-2025 | Brand Name: Dexcom ONE+ Continuous Glucose Monitoring System Product Name: Dexcom ONE+ iOS CGM App Model/Catalog Number: SW14244 Software Version: versions 1.4.0 and earlier Product Description: Dexcom ONE+ iOS CGM App Component: Dexcom ONE+ Continuous Glucose Monitoring System | 39309 | Class I | Ongoing | 10 Sep 2025 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class I: The highest of the three FDA recall classes.
The event in context
- Device events that started in 2025
- 945
- Share of these events in Class I
- 13%
- Product records in this event
- 5Median of the firm: 3
- Events of the firm on the register
- 11
Other events of the firm
Official channels
- FDA enforcement report Z-2446-2025The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 97262. https://recallregister.com/fda/i/97262/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
SW12299 G7 Android CGM App Worldwide Distribution - US Nationwide and the countries of Andorra, United Arab Emirates, Austria, Australia, Belgium, Bahrain, Canada, Switzerland, Czech Republic,…
- What quantity of product is in FDA event 97262?
- The event has 5 product records. The first record gives this quantity:
532821
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2025 have the same class?