Device recall: Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 US Retail Sensor/Transmitter 1-pack M
FDA enforcement event 98968 is a device recall by Dexcom, Inc. of San Diego, CA, initiated 27 May 2026 and classified Class II, with 1 product record.
FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 87% of the 572 device enforcement events initiated in 2026 carry the same class.
Status: Ongoing. Initiation: voluntary: firm initiated. Dexcom, Inc. has 11 enforcement events on this register, with a median of 3 product records per event.
Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.
Recalling firm
Dexcom, Inc., San Diego, CA.
Reason for recall
"Certain lots of Dexcom G7 sensors originally designated as scrap and intended for destruction were stolen during the destruction process. The sensors were later sold to pharmacies and distributors by Pharmsource, LLC. Lot 1725204004 were designated for destruction because the G7 sensors were not properly sterilized and have an increased risk of skin infection if used. Lot 1725069002 were…"
Distribution
"Dexcom does not have distribution records for this product due to its unauthorized release. However Dexcom is able to identify sensor session starts for the affected lots through the Dexcom G7 CGM…"
Product records
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-2574-2026 | Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 US Retail Sensor/Transmitter 1-pack Model/Catalog Number: STP-AT-012 Software Version: N/A Product Description: Dexcom G7 US Retail Sensor/Transmitter 1-pack Component: Dexcom G7 Continuous Glucose Monitoring System | N/A | Class II | Ongoing | 1 Jul 2026 |
Context
Other enforcement events for Dexcom, Inc.
All 11 events for Dexcom, Inc.
Questions and answers
Why was this device recalled?
FDA's record gives the reason for recall as: "Certain lots of Dexcom G7 sensors originally designated as scrap and intended for destruction were stolen during the destruction process. The sensors were later sold to pharmacies and distributors by Pharmsource, LLC. Lot 1725204004 were designated for destruction because the G7 sensors were not properly sterilized and have an increased risk of skin infection if used. Lot 1725069002 were…"
What does Class II mean?
FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.
Where was the product distributed?
FDA's distribution pattern field reads: "Dexcom does not have distribution records for this product due to its unauthorized release. However Dexcom is able to identify sensor session starts for the affected lots through the Dexcom G7 CGM…"
Is the recall still ongoing?