Recall Register

Device recall: Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 US Retail Sensor/Transmitter 1-pack…

Class II FDA enforcement event 98968. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

Certain lots of Dexcom G7 sensors originally designated as scrap and intended for destruction were stolen during the destruction process. The sensors were later sold to pharmacies and distributors by Pharmsource, LLC. Lot 1725204004 were designated for destruction because the G7 sensors were not properly sterilized and have an increased risk of skin infection if used. Lot 1725069002 were…

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Dexcom, Inc. started this device recall on 27 May 2026. The firm address in the FDA record is in San Diego, CA. FDA put the recall in Class II. FDA event 98968 has 1 product record. The first FDA enforcement report date is 1 Jul 2026. The record gives the status Ongoing. 291 of the 710 device events that started in 2026 have more product records. The median device event of 2026 has 1 product record.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Dexcom does not have distribution records for this product due to its unauthorized release. However Dexcom is able to identify sensor session starts for the affected lots through the Dexcom G7 CGM…

Recalling firm

Dexcom, Inc., San Diego, CA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 98968: product records
Recall numberProductQuantityClassStatusReport date
Z-2574-2026Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 US Retail Sensor/Transmitter 1-pack Model/Catalog Number: STP-AT-012 Software Version: N/A Product Description: Dexcom G7 US Retail Sensor/Transmitter 1-pack Component: Dexcom G7 Continuous Glucose Monitoring SystemN/AClass IIOngoing1 Jul 2026

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2026
710
Share of these events in Class II
87%
Product records in this event
1Median of the firm: 3
Events of the firm on the register
11

Other events of the firm

Other FDA enforcement events of Dexcom, Inc.
InitiatedProduct (first record)TypeClassRecords
14 Apr 2026Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM iOS App Model/Catalog Number: SW1DeviceClass II3
28 Aug 2025Brand Name: Dexcom G6 Continuous Glucose Monitoring System Product Name: Dexcom G6 and G6 Pro Android US CGM App Model/CDeviceClass I1
24 Jul 2025Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM Android App Model/Catalog Number:DeviceClass I5
12 May 2025Dexcom G6 Glucose Receiver, SKUs (Part Number): STK-FM-001 (MT27408-1) & STK-FR-001 (MT27408-1)DeviceClass I4
5 May 2025Dexcom G7 iOS Application, Ref: SW12300 used in conjunction with the Dexcom G7 Continuous Glucose Monitoring (CGM) systeDeviceClass II1
27 Jan 2025Dexcom G6, Continuous Glucose Monitoring System, REFs: STK-OM-001, STR-OM-001, STR-OR-001DeviceClass II3
3 Feb 2020Dexcom G4 PLATINUM Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is aDeviceClass II3
19 Nov 2019Brand Name: Dexcom G6 CGM App for iOS Software Number: SW11677 Software Version: Version 1.4.0 up to corrected Version 1DeviceClass II1
12 Jul 2019Dexcom Receiver, REF: MT22719, (Black, Pin, Blue), Rx Only, containing software version SW10617 and revision 4.0.1.048 -DeviceClass II1
23 Feb 2016Dexcom G4 PLATINUM Receiver Receiver Part Number: MT20649 Receiver (mg/dL), Black; MT20649-PNK, Receiver (mg/dL), Pink; DeviceClass I6

Dexcom, Inc.: all 11 events

Official channels

Cite this page

Recall Register. FDA enforcement event 98968. https://recallregister.com/fda/ii/98968/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
10 Nov 2026
Newest FDA report date on the register
4 Nov 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Dexcom does not have distribution records for this product due to its unauthorized release. However Dexcom is able to identify sensor session starts for the affected lots through the Dexcom G7 CGM…
What quantity of product is in FDA event 98968?
The FDA record gives this quantity: N/A
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2026 have the same class?
87% of the 710 device events that started in 2026 are in Class II.