Device recall: MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 13 cm, Curved Extensions, IC Tray REF:8888101001HP
Class I FDA enforcement event 92660. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow,…
Class I, in the words of FDA
a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death
Covidien LP started this device recall on 28 Jun 2023. The firm address in the FDA record is in Mansfield, MA. FDA put the recall in Class I. FDA event 92660 has 25 product records. The first FDA enforcement report date is 23 Aug 2023. The record gives the status Ongoing.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
US Nationwide - Worldwide Distribution Foreign: Argentina Austria Bahrain Bangladesh Belgium Brazil Brunei Darussalam Canada Cayman Islands China Costa Rica Czech Republic Denmark Dominican Republic…
Recalling firm
Covidien LP, Mansfield, MA.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
25 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-2367-2023 | MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 13 cm, Curved Extensions, IC Tray REF:8888101001HP | 2370 units | Class I | Ongoing | 23 Aug 2023 |
| Z-2368-2023 | MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 16 cm, Curved Extensions, IC Tray REF: 8888101002HP | 6615 | Class I | Ongoing | 23 Aug 2023 |
| Z-2369-2023 | MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 20 cm, Curved Extensions, IC Tray REF: 8888101003HP | 5373 units | Class I | Ongoing | 23 Aug 2023 |
| Z-2370-2023 | MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Curved Extensions, IC Tray REF: 8888101004HP | 1325 units | Class I | Ongoing | 23 Aug 2023 |
| Z-2371-2023 | MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 20 cm, Straight Extensions, IC Tray REF: 8888102003HP | 1875 units | Class I | Ongoing | 23 Aug 2023 |
| Z-2372-2023 | MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, IC Tray REF: 8888102004HP | 685 units | Class I | Ongoing | 23 Aug 2023 |
| Z-2373-2023 | MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 13 cm, Curved Extensions, PASS Tray REF: 8888103001HP | 299 units | Class I | Ongoing | 23 Aug 2023 |
| Z-2374-2023 | MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 20 cm, Curved Extensions, PASS Tray REF: 8888103002HP | 299 units | Class I | Ongoing | 23 Aug 2023 |
| Z-2375-2023 | MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Curved Extensions, PASS Tray REF: 8888103004HP | 271 units | Class I | Ongoing | 23 Aug 2023 |
| Z-2376-2023 | MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 20 cm, Straight Extensions, PASS Tray REF: 8888104003HP | 1099 units | Class I | Ongoing | 23 Aug 2023 |
| Z-2377-2023 | MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, PASS Tray REF: 8888104004HP | 519 units | Class I | Ongoing | 23 Aug 2023 |
| Z-2378-2023 | MAHURKAR 12 Fr Triple Lumen Acute Dialysis Catheter, 20 cm, Straight Extensions, Kit REF: 8888340629 | 655 units | Class I | Ongoing | 23 Aug 2023 |
| Z-2379-2023 | MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 20 cm, Straight Extensions, Kit REF: 8888340629HP | 7293 units | Class I | Ongoing | 23 Aug 2023 |
| Z-2380-2023 | MAHURKAR 12 Fr Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, Kit REF: 8888340637 | 775 units | Class I | Ongoing | 23 Aug 2023 |
| Z-2381-2023 | MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, Kit REF: 8888340637HP | 2425 units | Class I | Ongoing | 23 Aug 2023 |
| Z-2382-2023 | MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 16 cm, Curved Extensions REF: 8888345512HP | 600 units | Class I | Ongoing | 23 Aug 2023 |
| Z-2383-2023 | MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 20 cm, Curved Extensions REF: 8888345520HP | 295 units | Class I | Ongoing | 23 Aug 2023 |
| Z-2384-2023 | MAHURKAR 12 Fr Triple Lumen Acute Dialysis Catheter, 13 cm, Curved Extensions, Kit" REF: 8888345603 | 45880 units | Class I | Ongoing | 23 Aug 2023 |
| Z-2385-2023 | MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 13 cm, Curved Extensions, Kit REF: 8888345603HP | 2685 units | Class I | Ongoing | 23 Aug 2023 |
| Z-2386-2023 | MAHURKAR 12 Fr Triple Lumen Acute Dialysis Catheter, 16 cm, Curved Extensions, Kit" REF: 8888345611 | 69987 units | Class I | Ongoing | 23 Aug 2023 |
| Z-2387-2023 | MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 16 cm, Curved Extensions, Kit REF: 8888345611HP | 32125 units | Class I | Ongoing | 23 Aug 2023 |
| Z-2388-2023 | MAHURKAR 12 Fr Triple Lumen Acute Dialysis Catheter, 20 cm, Curved Extensions, Kit" REF: 8888345629 | 6949 units | Class I | Ongoing | 23 Aug 2023 |
| Z-2389-2023 | MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 20 cm, Curved Extensions, Kit REF: 8888345629HP | 15209 units | Class I | Ongoing | 23 Aug 2023 |
| Z-2390-2023 | MAHURKAR 12 Fr Triple Lumen Acute Dialysis Catheter, 24 cm, Curved Extensions, Kit REF: 8888345637 | 647 units | Class I | Ongoing | 23 Aug 2023 |
| Z-2391-2023 | MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Curved Extensions, Kit REF: 8888345637HP | 3965 units | Class I | Ongoing | 23 Aug 2023 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class I: The highest of the three FDA recall classes.
The event in context
- Device events that started in 2023
- 917
- Share of these events in Class I
- 10%
- Product records in this event
- 25Median of the firm: 2
- Events of the firm on the register
- 31
Other events of the firm
Official channels
- FDA enforcement report Z-2367-2023The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 92660. https://recallregister.com/fda/i/92660/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
US Nationwide - Worldwide Distribution Foreign: Argentina Austria Bahrain Bangladesh Belgium Brazil Brunei Darussalam Canada Cayman Islands China Costa Rica Czech Republic Denmark Dominican Republic…
- What quantity of product is in FDA event 92660?
- The event has 25 product records. The first record gives this quantity:
2370 units
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2023 have the same class?
- 10% of the 917 device events that started in 2023 are in Class I.