Food, drugs, devices › Philips Respironics, Inc.

Device recall: Trilogy Evo, Garbin Evo, and Aeris EVO, Model Numbers DS2110X11B, IN2110X15B, JP2110X16B, AU2110X15B, LA2110X15B, CA2110

FDA enforcement event 92048 · Class I · initiated 1 May 2023 · status Ongoing

DeviceVoluntary: Firm initiated4 product recordsMurrysville, PA

Philips Respironics, Inc. (Murrysville, PA) initiated this device recall on 1 May 2023; FDA lists 4 product records under event 92048, classified Class I.

FDA defines Class I as a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death. 10% of the 917 device enforcement events initiated in 2023 carry the same class.

Status: Ongoing. Initiation: voluntary: firm initiated. Philips Respironics, Inc. has 26 enforcement events on this register, with a median of 3 product records per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Philips Respironics, Inc., Murrysville, PA.

Reason for recall

"Environmental debris may accumulate on the internal machine flow sensor causing partial occlusion which may impact accurate delivery of pressure, volume, or flow. A malfunction caused by this issue can result in patient harm up to hypoxemia if not addressed by the care provider."

Distribution

"US Nationwide. Global Distribution."

Product records

Recall numberProductQuantityClassStatusReport date
Z-1649-2023Trilogy Evo, Garbin Evo, and Aeris EVO, Model Numbers DS2110X11B, IN2110X15B, JP2110X16B, AU2110X15B, LA2110X15B, CA2110X12B, BR2110X18B, KR2110X15B, IA2110X15B, FP2110X10, FR2110X14B, ND2110X15B, IT2110X21B, ES2110X15B, DE2110X13B, BL2110X15B, GB2110X15B, EU2110X15B, EE2110X15B, LD2110X23B, and VT2110X24B57,395 devicesClass IOngoing14 Jun 2023
Z-1650-2023Trilogy Evo O2 and LifeVent EVO2, Model Numbers DS2100X11B, IN2100X15B, IN2100X19, FX2100X15B, LA2100X15B, IA2100X15B, FP2100X10, FR2100X14B, ND2100X15B, IT2100X21B, ES2100X15B, DE2100X13B, EU2100X15B, EU2100X19, EE2100X15B, and SP2100X26B21,520 devicesClass IOngoing14 Jun 2023
Z-1651-2023Trilogy Evo O2 EV300, Model Numbers DS2200X11B, IN2200X15B, FX2200X15B, CA2200X12B, FR2200X14B, ND2200X15B, ES2200X15B, BL2200X15B, GB2200X15B, EU2200X15B, EE2200X15B, and TR2200X15B18,855 devicesClass IOngoing14 Jun 2023
Z-1652-2023Trilogy Evo Universal Ventilator, Model Number DS2000X11B22,237 devicesClass IOngoing14 Jun 2023

Context

10%of 2023 device events are Class I917 events that year
4product records in this eventfirm median: 3
26events on this register for the firm

Other enforcement events for Philips Respironics, Inc.

InitiatedProduct (first record)TypeClassRecords
2 Mar 2026Philips Respironics Trilogy Evo, Software Version 1.05.15.00. Continuous home-use ventilator device.DeviceClass I4
2 Mar 2026Philips Respironics Trilogy Evo, Software Version 1.05.15.00. Continuous home-use ventilator device.DeviceClass I4
7 Oct 2025SimplyGo Mini, Standard Battery Kit; Product code: 1116816; The affected battery kits are sold as accessories for the SiDeviceClass II1
30 Jun 2025DreamStation Auto CPAP. Non-Continuous Ventilator.DeviceClass I3
9 May 2025BiPAP V30 Auto Ventilator. Intended to provide non-invasive ventilatory support for pediatric and adult patients weighinDeviceClass I3
7 Oct 2024Trilogy Evo O2. Model Numbers: FP2100X10, DS2100X11B, IN2100X15B, DE2100X13B, IA2100X15B, EU2100X15B, LA2100X15B, EE2100DeviceClass I7
16 Jul 2024Trilogy Evo EV300 -- Model Number/ Model Description: BL2200X15B Trilogy Evo, O2, Benelux EV300; BR2200X18B Trilogy Evo,DeviceClass I3
1 Apr 2024OmniLab Advanced + Ventilator Product Numbers 1111122 1111123 1111124 1111125 1111126 1111127 1111138 1111141 1111142 KRDeviceClass I1
26 Mar 2024BiPAP V30 Auto Ventilator, Part Number 1111178DeviceClass I3
19 Mar 2024Aeris EVO, all software versions except 1.05.06.00. Ventilator for pediatric and adult patients. (sold outside US)DeviceClass I7

All 26 events for Philips Respironics, Inc.

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "Environmental debris may accumulate on the internal machine flow sensor causing partial occlusion which may impact accurate delivery of pressure, volume, or flow. A malfunction caused by this issue can result in patient harm up to hypoxemia if not addressed by the care provider."

What does Class I mean?

FDA classifies recalls by relative degree of health hazard. Class I is a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death.

Where was the product distributed?

FDA's distribution pattern field reads: "US Nationwide. Global Distribution."

Is the recall still ongoing?

FDA lists the status as Ongoing. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.