Recall Register

Device recall: Trilogy Evo O2 and LifeVentEVO2 Portable Electric Ventilator, Model Numbers DS2100X11B, IN2100X15B, IN2100X19…

Class I FDA enforcement event 91832. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

The accuracy of delivered oxygen may deviate below the required tolerance of 5% from setpoint when providing high concentration oxygen therapy. Additionally, if equipped, the internal FiO2 sensor may indicate a value higher than the device is actually delivering. This may lead to under delivery of oxygen.

Class I, in the words of FDA

a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death

Philips Respironics, Inc. started this device recall on 1 Mar 2023. The firm address in the FDA record is in Murrysville, PA. FDA put the recall in Class I. FDA event 91832 has 3 product records. The first FDA enforcement report date is 12 Apr 2023. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide distribution. US Nationwide including Puerto Rico, and countries of AE, AR, AT, AU, BD, BE, BG, BO, BR, CA, CH, CL, CN, CO, CZ, DE, DK, EC, EE, EG, ES, FI, FR, GB, GE, GT, HK, HR, HU, ID,…

Recalling firm

Philips Respironics, Inc., Murrysville, PA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

3 product records in the event.

FDA event 91832: product records
Recall numberProductQuantityClassStatusReport date
Z-1287-2023Trilogy Evo O2 and LifeVentEVO2 Portable Electric Ventilator, Model Numbers DS2100X11B, IN2100X15B, IN2100X19, FX2100X15B, JP2100X16B, AU2100X15B, LA2100X15B, CA2100X12B, CN2100X17B, BR2100X18B, KR2100X15B, IA2100X15B, PP2100X10, FP2100X10, FR2100X14B, ND2100X15B, IT2100X21B, ES2100X15B, DE2100X13B, BL2100X15B,…62,826 total devicesClass IOngoing12 Apr 2023
Z-1288-2023Trilogy EV300 Portable Electric Ventilator, Model Numbers DS2200X11B, IN2200X15B, FX2200X15B, CA2200X12B, CN2200X17B, BR2200X18B, KR2200X15B, IA2200X15B, FR2200X14B, ND2200X15B, IT2200X21B, ES2200X15B, DE2200X13B, BL2200X15B, GB2200X15B, EU2200X15B, EU2200X19, EE2200X15B, TR2200X15B, RU2200X15B62826 total devicesClass IOngoing12 Apr 2023
Z-1289-2023Trilogy Evo Universal Portable Electric Ventilator, Model Number DS2000X11B62826 total devicesClass IOngoing12 Apr 2023

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class I: The highest of the three FDA recall classes.

The event in context

Device events that started in 2023
917
Share of these events in Class I
10%
Product records in this event
3Median of the firm: 3
Events of the firm on the register
26

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 91832. https://recallregister.com/fda/i/91832/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide distribution. US Nationwide including Puerto Rico, and countries of AE, AR, AT, AU, BD, BE, BG, BO, BR, CA, CH, CL, CN, CO, CZ, DE, DK, EC, EE, EG, ES, FI, FR, GB, GE, GT, HK, HR, HU, ID,…
What quantity of product is in FDA event 91832?
The event has 3 product records. The first record gives this quantity: 62,826 total devices
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2023 have the same class?
10% of the 917 device events that started in 2023 are in Class I.