Device recall: HAMILTON-C6 Intensive Care Ventilator, REF: 160021
Class I FDA enforcement event 90576. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
Ventilator status indicator board can become loose, which could lead to water ingress (disinfectants) that may lead to technical fault alarms. Multiple technical faults in a short time may force ventilators into Safety Ventilation (blower runs constantly) or Ambient Sate (inspiratory channel/expiratory valves opened; patient breaths room air unassisted) with Panel connection lost message displayed
Class I, in the words of FDA
a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death
Hamilton Medical AG started this device recall on 27 Jun 2022. The firm address in the FDA record is in Bonaduz. FDA put the recall in Class I. FDA event 90576 has 1 product record. The first FDA enforcement report date is 31 Aug 2022. The record gives the status Ongoing. 254 of the 839 device events that started in 2022 have more product records. The median device event of 2022 has 1 product record. The openFDA file has 839 device events that started in 2022 and 900 that started in 2021.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
US nationwide distribution including Puerto Rico.
Recalling firm
Hamilton Medical AG, Bonaduz, Switzerland.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
1 product record in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1525-2022 | HAMILTON-C6 Intensive Care Ventilator, REF: 160021 | 358 | Class I | Ongoing | 31 Aug 2022 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class I: The highest of the three FDA recall classes.
The event in context
- Device events that started in 2022
- 839
- Share of these events in Class I
- 8%
- Product records in this event
- 1Median of the firm: 1
- Events of the firm on the register
- 12
Other events of the firm
Official channels
- FDA enforcement report Z-1525-2022The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 90576. https://recallregister.com/fda/i/90576/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 10 Nov 2026
- Newest FDA report date on the register
- 4 Nov 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
US nationwide distribution including Puerto Rico.
- What quantity of product is in FDA event 90576?
- The FDA record gives this quantity:
358
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2022 have the same class?
- 8% of the 839 device events that started in 2022 are in Class I.