Device recall: Medtronic Bio-Console 560 Extracorporeal Blood Pumping Console, Model Numbers: 560BC, 560BC1, 560BCS, 560BCS1…
Class I FDA enforcement event 87490. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
An electrical component within the system controller module of certain Bio-Console 560 devices may exhibit electrical failure causing the device to stop pumping. Issues observed include the pump stopping, the user interface going blank, and smoking and producing a burning odor during use.
Class I, in the words of FDA
a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death
Medtronic Perfusion Systems started this device recall on 3 Mar 2021. The firm address in the FDA record is in Brooklyn Park, MN. FDA put the recall in Class I. FDA event 87490 has 1 product record. The first FDA enforcement report date is 7 Apr 2021. The record gives the status Ongoing.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide distribution. US Nationwide including Puerto Rico, Brazil, Brunei Darussalam, Canada, China, Colombia, Croatia, Ecuador, Guatemala, India, Israel, Italy, Japan, Kazakhstan, Korea, Mexico,…
Recalling firm
Medtronic Perfusion Systems, Brooklyn Park, MN.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
1 product record in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1313-2021 | Medtronic Bio-Console 560 Extracorporeal Blood Pumping Console, Model Numbers: 560BC, 560BC1, 560BCS, 560BCS1, R560BCS1. Used in cardiopulmonary bypass (CPB) procedures. | 387 devices (93 US) | Class I | Ongoing | 7 Apr 2021 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class I: The highest of the three FDA recall classes.
The event in context
- Device events that started in 2021
- 900
- Share of these events in Class I
- 8%
- Product records in this event
- 1Median of the firm: 1
- Events of the firm on the register
- 38
Other events of the firm
Official channels
- FDA enforcement report Z-1313-2021The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 87490. https://recallregister.com/fda/i/87490/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide distribution. US Nationwide including Puerto Rico, Brazil, Brunei Darussalam, Canada, China, Colombia, Croatia, Ecuador, Guatemala, India, Israel, Italy, Japan, Kazakhstan, Korea, Mexico,…
- What quantity of product is in FDA event 87490?
- The FDA record gives this quantity:
387 devices (93 US)
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2021 have the same class?
- 8% of the 900 device events that started in 2021 are in Class I.