Device recall: Aortic Root Cannula: 1. DLP Aortic Root Cannula A. CANNULA 11012 AR FLOWGUARD 12GA, model no 11012 B. CANNULA 11014 AR…
Class I FDA enforcement event 96259. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
Unexpected loose material in the male luer used in the aortic root cannula has been identified. Potential harms when identified prior to use - procedure delay. Potential harm if not identified and used - stroke (reversible and irreversible).
Class I, in the words of FDA
a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death
Medtronic Perfusion Systems started this device recall on 5 Feb 2025. The firm address in the FDA record is in Brooklyn Park, MN. FDA put the recall in Class I. FDA event 96259 has 1 product record. The first FDA enforcement report date is 19 Mar 2025. The record gives the status Ongoing. 352 of the 945 device events that started in 2025 have more product records. The median device event of 2025 has 1 product record. The openFDA file has 945 device events that started in 2025 and 972 that started in 2024.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Domestic distribution nationwide. International distribution to the following countries: Albania Algeria Argentina Armenia Australia Austria Azerbaijan Bahrain Bangladesh Belarus Belgium Bosnia And…
Recalling firm
Medtronic Perfusion Systems, Brooklyn Park, MN.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
1 product record in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1306-2025 | Aortic Root Cannula: 1. DLP Aortic Root Cannula A. CANNULA 11012 AR FLOWGUARD 12GA, model no 11012 B. CANNULA 11014 AR FLOWGUARD 14GA, model no. 11014 2. MiAR Cannula: A. CANNULA 11012L AR FLOWGUARD LONG 12GA, model no 11012L B. CANNULA 11014L AR FLOWGUARD 14FR LONG, model no. 11014L 3. DLP Aortic Root Cannula with… | 316,205 | Class I | Ongoing | 19 Mar 2025 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class I: The highest of the three FDA recall classes.
The event in context
- Device events that started in 2025
- 945
- Share of these events in Class I
- 13%
- Product records in this event
- 1Median of the firm: 1
- Events of the firm on the register
- 38
Other events of the firm
Official channels
- FDA enforcement report Z-1306-2025The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 96259. https://recallregister.com/fda/i/96259/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 10 Nov 2026
- Newest FDA report date on the register
- 4 Nov 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Domestic distribution nationwide. International distribution to the following countries: Albania Algeria Argentina Armenia Australia Austria Azerbaijan Bahrain Bangladesh Belarus Belgium Bosnia And…
- What quantity of product is in FDA event 96259?
- The FDA record gives this quantity:
316,205
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2025 have the same class?
- 13% of the 945 device events that started in 2025 are in Class I.