Device recall: IntraClude Intra-Aortic Occlusion Device, REF: ICF100, 10.5 Fr (3.5 mm) X 39.4" (100 cm), Rx Only, Sterile EO The…
Class I FDA enforcement event 82870. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
Possibility of balloon rupture in intra-aortic occlusion devices. If the balloon bursts during cardiopulmonary bypass, the heart can fill and warm, the operative site may be obscured, and the device will need to be exchanged or operative strategy would need to change, including placement of an external cross-clamp, conversion to an open procedure, or performing the procedure under fibrillation.
Class I, in the words of FDA
a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death
Edwards Lifesciences, LLC started this device recall on 14 May 2019. The firm address in the FDA record is in Irvine, CA. FDA put the recall in Class I. FDA event 82870 has 1 product record. The first FDA enforcement report date is 19 Jun 2019. The record gives the status Terminated, with the termination date 23 Apr 2024.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide Distribution: US (nationwide) to states of: UT, GA, NJ, CA, OH, PA, AL, FL, TN, TX, ME, MA, IN, MS, NY, VA, MI, IL, CO, OR, WA, SC, AZ, SD, MO, CT, OK, NV, WV, WI, NC, MN, KY; and…
Recalling firm
Edwards Lifesciences, LLC, Irvine, CA.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
1 product record in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1705-2019 | IntraClude Intra-Aortic Occlusion Device, REF: ICF100, 10.5 Fr (3.5 mm) X 39.4" (100 cm), Rx Only, Sterile EO The IntraClude intra-aortic occlusion device is indicated for use in patients undergoing cardiopulmonary bypass. The IntraClude intra-aortic occlusion device occludes and vents the ascending aorta when the… | 6552 | Class I | Terminated | 19 Jun 2019 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class I: The highest of the three FDA recall classes.
The event in context
- Device events that started in 2019
- 968
- Share of these events in Class I
- 6%
- Product records in this event
- 1Median of the firm: 1
- Events of the firm on the register
- 50
Other events of the firm
Official channels
- FDA enforcement report Z-1705-2019The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 82870. https://recallregister.com/fda/i/82870/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide Distribution: US (nationwide) to states of: UT, GA, NJ, CA, OH, PA, AL, FL, TN, TX, ME, MA, IN, MS, NY, VA, MI, IL, CO, OR, WA, SC, AZ, SD, MO, CT, OK, NV, WV, WI, NC, MN, KY; and…
- What quantity of product is in FDA event 82870?
- The FDA record gives this quantity:
6552
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2019 have the same class?
- 6% of the 968 device events that started in 2019 are in Class I.